xxii
●
Compliance to the Electromagnetic Emissions
The DS-7100 system is intended for use in the electromagnetic environment specified below.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1
The equipment uses RF energy that is necessary for the internal
functioning of the equipment itself. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby
electronic equipment.
RF Emissions
CISPR 11
Class A
This equipment is suitable for use in all establishments other than
domestic and those directly connected to a low-voltage power
supply network which supplies buildings used for domestic
purposes.
●
Compliance to the Electromagnetic Immunity (1)
The DS-7100 system is intended for use in the electromagnetic environment specified below. It
should be assured that the DS-7100 system is used in such an environment.
Immunity Test
IEC60601-1-2
Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic
Discharge (ESD)
IEC61000-4-2
?
3kV contact
?
8kV air
?
3kV contact
?
8kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient / burst
IEC61000-4-4
?
1kV for power supply
lines
?
0.5kV for input/output
lines
?
1kV for power supply
lines
?
0.5kV input/output
lines
Power supply quality should be at
levels characteristic of a typical
location in a typical commercial or
hospital environment.
Surge
IEC61000-4-5
?
2kV:
line to ground
?
1kV:
line to line
?
2kV:
line to ground
?
1kV:
line to line
Power supply quality should be at
levels characteristic of a typical
location in a typical commercial or
hospital environment.