xx
C A U T I O N
?
Do not peel off or scratch the exterior tubing.
?
Do not use or leave the battery in high temperature. It may result in
leakage or deterioration of the performance / life cycle of the battery.
?
Immediately stop using the battery if any abnormality is found during use.
?
Do not use / store the battery in reach of infants.
?
If the monitor is expected not to be used for a long time, turn off the power
and unplug the power cable from the outlet. Otherwise it may result in
battery leakage.
N O T E
Users should not attempt to install or replace the battery pack.
For installation and replacement of the battery pack, contact our service
representative.
To Prepare for Emergency Use
1.
Battery Pack
(1) The battery self-discharges even when not in use. If there is any possibility to use the
battery in emergency, the power cable should be always connected to the power receptacle.
To fully charge the empty battery, it takes approximately 2.5 hours when the monitor is not
operating, and approximately 13 hours when the monitor is operating.
Reference
Refer to “2. Basic Operation To Use with the Battery Pack”
(2) The performance of the battery deteriorates with the repeated use. To maintain the initial
performance, replace the battery at least once a year. It is recommended to indicate the
start usage date on the battery so that replacing date can be easily recognized.
2.
Accessories / Optional Accessories
(1) The ECG electrodes are consumables. Always prepare extra supplies of electrodes.
(2) Check if any wire break on the patient cables once a week.