Precautions for Use of SpO
2
Sensor
D A N G E R
Burn Risk in Using SpO
2
Sensor
In SpO
2
monitoring, always use the sensor/relay cable specified by Fukuda
Denshi. If any other sensor/relay cable is used, a high temperature rise of
the sensor may place the patient in danger of burns.
If there are any questions regarding the sensor/relay cable use for SpO
2
measurements of this device, please contact Fukuda Denshi service
representative.
Precautions for Use of NIBP Cuff
C A U T I O N
This product contains natural rubber latex which may cause allergic reactions.
Disposing of Equipment, Accessories, or Components
C A U T I O N
When disposing of the equipment, accessories, or components, use an
industrial waste distributor. Do not dispose of as ordinary waste.
Precautions about Transportation
For transporting the DS-7100 system, pack with specified packing materials.
Reference
Refer to “11. Technical Information Specification / Performance” for environmental condition
during transportation.
Precautions about RTC or Data Backup
C A U T I O N
?
The DS-7100 system is equipped with a built-in clock. When the power of
the DS-7100 system is turned off, this clock is backed up by a lithium primary
battery.
If incorrect time is displayed when turning on the power, a low battery may be
the cause. In such case, contact Fukuda Denshi service representative for
replacing the battery.
?
To protect the data during voltage dip, short interruptions and voltage
variations on power supply input lines or during short duration of power
turned OFF, this monitor performs 5-minute (approx.) data backup using the
secondary battery. The data may not be protected if the power is turned off
within 30 minutes from power on.