Pulse Oximeter
Measurement Update Rate: 1 sec.
Nellcor Unit (HS-8312N)
Measurement Method : 2 Wavelength Pulse Wave Method
Wavelength: Approx. 660nm (red light)
Approx. 890nm (infrared light)
Output: 15mW or below
Measurement Range
: 1 to 100%
Resolution :
1%
Measurement Accuracy : Adult: ±3% when 70 to 100% (DS-100A)
Neonate: ±2% when 70 to 100%
PR Measurement Range: 20 to 250bpm
PR Resolution
: 1bpm
PR Accuracy
: ±3bpm when 20 to 250bpm
Measurement Response: 6 to 7 sec. (averaging duration)
Time
Masimo Unit (HS-8312M)
Measurement Method : 2 Wavelength Pulse Wave Method
Masimo LNOP/LNCS Sensor
Wavelength: Approx. 660nm (red light)
Approx. 905nm (infrared light)
Output: 15mW or below
Masimo Rainbow Sensor
Wavelength: 8 different wavelengths are used within the range of
610nm to 905nm
Output: 25mW or below
SpO
2
Measurement Range : 1 to 100%
Resolution :
1%
Measurement Accuracy: Adult: ±2% when 70 to 100%
Neonate: ±3% when 70 to 100%
PI (Perfusion Index)
Measurement Range : 0.02 to 20%
PVI
Measurement Range : 0 to 100%
SpCO
Measurement Range : 0 to 99%
Resolution :
1%
Measurement Accuracy: ±3% when 1 to 40%
SpMet
Measurement Range : 0 to 99.9%
Resolution :
0.1%
Measurement Accuracy: ±1% when 1 to 15%
SpHb
Measurement Range : 0 to 25.0 g/dL
Resolution :
0.1g/dL
Measurement Accuracy: ±1g/dL when 8 to 17 g/dL
Pulse Rate
Measurement Range : 26 to 239bpm
Measurement Accuracy: ±3bpm when 26 to 239bpm (without body motion)
Measurement Response : 7 levels
Time
2 to 4 sec., 4 to 6 sec., 8 sec., 10 sec., 12 sec., 14 sec., 16 sec.
(averaging
duration)
2-10