No.
Check Item
Check Procedure
Criteria
14. Electrical Safety (*)
Perform the following check items as appropriate. Check these items when you have disassembled the equipment to
check / replace the boards or units.
08 Protective
Ground
Resistance
Measure the protective ground
continuity resistance.
Comply with the test method of IEC 60601-1 18.
Should be 0.1
Ω
or below
between the protective
ground terminal and
accessible conductive part.
Protective ground
resistance in the power
cable should be 0.1
Ω
or
below.
09 Withstand
Voltage
Test (A-a1)
Apply AC1500V for 1 minute between
primary power source and accessible
metal part with protective ground
(A-a1).
Comply with the test method of IEC 60601-1 20.
Should withstand applied
voltage
10 Withstand
Voltage
Test (A-a2)
Apply AC4000V for 1 minute between
primary power source and exterior
without protective ground (A-a2).
Comply with the test method of IEC 60601-1 20.
Should withstand applied
voltage
11 Withstand
Voltage
Test (B-a)
Apply AC4000V for 1 minute between
primary power source and applied part
(B-a).
Note: Since the internal protective
circuit configuration differs
depending on the device, test
voltage may differ.
Comply with the test method of IEC 60601-1 20.
Should withstand applied
voltage
12 Withstand
Voltage
Test (B-d)
Apply AC1500V for 1 minute between
the exterior and applied part (B-d).
Note: Since the internal protective
circuit configuration differs
depending on the device, test
voltage may differ.
Comply with the test method of IEC 60601-1 20.
Should withstand applied
voltage.
11-13