TN5A1191-E
23
3.4 Calibration coefficient
z
This menu allows you to change the coefficient for zero calibration and for span calibration.
Factory mode > 14. Coefficient
Factory mode initial screen
The cursor is in 14.
ENT
ESC
- “R1 and R2” means the range 1 and the range 2
respectively.
- The third column from left shows the zero calibration
coefficient, and the fourth shows the span coefficient.
ENT
ESC
Select the component and range you want edit, and
press the
ENT
key. Use the
and key to change
the value, and the
key to select the digit, and press
the
ENT
key. If you want to cancel the change, press the
ESC
key before pressing the
ENT
key.