TN5A1191-E
21
3.2 Zero limit
z
You can switch whether or not to show the values below zero of the measured value.
OFF: Values below zero are not displayed
ON: Values below zero are displayed
z
The default setting is “OFF”.
If you turn “ON” the zero limit to check the indication during analyzer adjustment, do not forget to turn
it “OFF” after adjustment.
Factory mode > 12. Other parameters > Zero limit
Factory mode initial screen
The cursor is in 12.
ENT
ESC
Point the cursor to “Zero limit” and press the
ENT
key.
The setpoint (ON or OFF) is inverted.
ENT
ESC
By using the
or key, switch between ON and
OFF, and press the
ENT
key. If you want to cancel the
change, press the
ESC
key before pressing the
ENT
key.