Service Manual for FNDX 5
08/2009 FR
Page 9
7.1.1
THE FIRST STEPS
WARNING:
Separate the Film Processor from mains. To do so, switch the main power switch
of the Filmprocessor to "0" position.
Wear safety goggles, protection gloves and clothing!
7.1.2
ADDITIONAL TO THE FIRST STEPS
We recommend that the machine is fully water tested on installation before filling with chemistry – this is
just a safety procedure in case of transport damage.
*)
Thoroughly clean processor ensuring no packing materials restricts the free running ofthe processor.
Pay special attention to the racks and inside of the tanks.
*)
Close the drain taps for the developer, fixer and wash.
*)
Fill the tanks (Dev, Fix and Wash) with water to the markers on the tank wall.
*)
Switch on the processor
*)
Check for signs of leakage.
7.1.3
USING THE CHEMISTRIES
*)
Only use chemistry suitable for roller transport systems.
*)
Follow instructions of chemistry manufacturers.
FIXER BATH:
*)
Empty fixer tank by opening the fix drain tap.
*)
Remove the Fixer-rack.
*)
Check fixer tank is free of alien material.
*)
Close fix drain tap.
*)
Fill fixer tank with ready-to-use-fixer solution to the red marker on the tank wall.
Insert the Fixer-rack very carefully and slowly, add hardener solution if advised by the chemistry
manufacturer.
DEVELOPER BATH:
*)
Empty developer tank by opening dev drain tap.
*)
Remove the Developer-rack.
*)
Check developer tank is free of alien material.
*)
Close dev drain tap.
*)
Fill developer tank with ready-to-use-developer solution to the red marker on the tank wall.
Insert the developer-rack very carefully and slowly. Replenishment tanks may be used to mix the
chemistry. Any remaining can be used for replenishment.CAUTION:
Even the smallest quantity of fixer could contaminate the developer solution. Therefore, always
fill with fixer first.
When removing the fixer rack, always cover the developer tank.
For removing the fixer rack use rack carrier tray (optional accessory).
CAUTION:
Even the smallest quantity of fixer could contaminate the developer
solution. Therefore, always fill with fixer first.
When removing the fixer rack, always cover the developer tank.
For removing the fixer rack use rack carrier tray (optional accessory).