73
INZ-TN2ZPAF-E
EN61000-3-3:2008
Electrical equipment for measurement, control
and laboratory use — EMC requirements.
6. Requirements for Sample Gas
Flow rate:
0.5 ±0.2L / min (including purge gas for
H
2
S measurement)
7HPSHUDWXUH
WRÛ&
Pressure:
10 kPa or less (Gas outlet side should be
open to the atmospheric air.)
Dust: 100
μ
g/Nm
3
or less in particle size of 0.3
μ
m or smaller
Mist: Unallowable
Moisture:
Less than the content saturated at 2°C
Corrosive component: 1 ppm or less
(H
2
S scrubber is required on pipings for
NDIR and O
2
measurement.)
Standard gas for calibration:
1) Infrared-ray measurable component, standard O
2
Zero gas ; Dry air
Span gas ; Each sample gas having
concentration 90 to 100% of
its measuring range (recom-
mended).
2)
H
2
S measurement
Zero gas: air *
Span gas: concentration of 90 to 100 %
of its measuring range
Purge gas: air *
*Use moist air saturated at the temperature
from room temperature through 2°C. Do
not use air which includes H2S nor dry
air saturated below 2°C or less.
7. Installation Requirements
,QGRRU XVH 6HOHFW D SODFH ZKHUH WKH HTXLSPHQW GRHV
not receive direct sunlight, draft/rain or radiation from hot
substances. If such a place cannot be found, a roof or
cover should be prepared for protection.)
$YRLGDSODFHZKHUHXQLWUHFHLYHVKHDY\YLEUDWLRQ
6HOHFWDSODFHZKHUHDWPRVSKHULFDLULVFOHDQ
$QDO\]HUSXUJHZLWK1
2
or air is indispensable.
+DQGOH+
2
6ZLWKJUHDWFDUHDVLWLVWR[LFÀDPPDEOHDQG
corrosive.
)RUVDIHW\LQ
2
S alarm around the analyzer.
8. Items to be Prepared Saparately
&RQVWDQWSRWHQWLDOHOHFWURO\WLF+
2
S sensor (for replacement):
TQ503691C1
*DOYDQLF2
2
sensor (for replacement): TQ503691C2