21
INZ-TN2ZPAF-E
6. SETTING
AND CALIBRATION
6.1 Switch of range
6.1.1 Setting of range switch mode
6HWWKHUDQJHVZLWFKPRGHDVIROORZV
3UHVVWKH
MODE
NH\LQPHDVXUHPHQW
PRGHWRGLVSOD\WKH8VHUPRGHVFUHHQ
0RYHWKHFXUVRUWR³6ZLWFK5DQJHV´DQG
SUHVVWKH
ENT
key.
,QWKH³&KDQQHO6HOHFWLRQ´VFUHHQWKDWDS
SHDUVPRYHWKH FXUVRUE\SUHVVLQJWKH
RUWKH NH\DQGVHOHFW&KFRPSR
QHQW
7KHQSUHVVWKH
ENT
key.
6HOHFWHGUDQJHVZLWFKPRGHLVKLJKOLJKW
ed.
3UHVVWKH RUWKH NH\WRVHOHFWDGH
VLUHGVZLWFKPRGH
7KHQSUHVVWKH
ENT
NH\WRFRQ¿UPWKH
VHOHFWLRQ
,I³05´LVVHOHFWHGWKHFXUVRUPRYHVWR
³5DQJH6ZLWFK´
MR: Select a desired range on this screen.
RR: Select a desired range according to
the remote range switch contact input.
AR: Automatically switched from Range 1
to Range 2 when the measured concen-
tration exceeds 90% of Range 1.
Automatically switched from Range 2
to Range 1 when the measured concen-
tration becomes less than 80% of
Range
1.
* Operation set for each Ch only can be
performed.
Description of setting
( )
ENT
MODE
ENT
ENT
Range switch
or previous screen
User Mode
Switch Ranges
Calibration Parameters
Alarm Setting
Setting of Auto Calibration
Setting of Auto Zero Calibration
Parameter Setting
Select an item
with UP/DOWN and ENT
Back with ESC
Switch Range
Range1
Range2
0–20.00
0–100.0
vol%
vol%
Range1
Range2
0–20.00
0–100.0
vol%
vol%
Range1
Range2
0–500.0
0–2000
ppm
ppm
Ch1
CH
4
Ch2
CO
2
H
2
S
Ch3
AR
RR
Range1
Range2
0–10.00
0–25.00
vol%
vol%
O
2
Ch4
MR
Select Ch No.
with UP/DOWN and ENT
Back with ESC
MR
Switch Range
Range1
Range2
0–20.00
0–100.0
vol%
vol%
Range1
Range2
0–20.00
0–100.0
vol%
vol%
Range1
Range2
0–500.0
0–2000
ppm
ppm
Ch1
CH
4
Ch2
CO
2
H
2
S
Ch3
AR
RR
Range1
Range2
0–10.00
0–25.00
vol%
vol%
O
2
Ch4
MR
Select Ch No.
Switch ranges
with UP/DOWN and ENT
Back with ESC
MR