- 87 -
INZ-TN2ZKMA-E
Hysteresis
:
If the value fluctuates around the condition value, there is a possibility that alarms occur frequent-
ly. When determining alarms, set a hysteresis width for the condition in order to prevent chattering.
For alarm check, set the percentage (%) of the range compared to the full scale as hysteresis width
(see the figure below).
This is common among “HH limit value,” “High limit value,” “Lower limit value,” and “LL limit
value.”
High limit judgment
Lower limit judgment
High limit value
Lower limit value
Hysteresis
Oxygen concentration value
HH limit value
LL limit judgment
HH limit judgment
LL limit value
HH / LL limit judgment
High / lower limit judgment
Summary of Contents for ZKMA
Page 1: ......
Page 13: ...2 INZ TN2ZKMA E 2 OPERATING PARTS AND THEIR FUNCTIONS 2 1 Outline Drawing 1 ZKMA 2 ZKMB...
Page 22: ...11 INZ TN2ZKMA E 4 2 2 Cable gland and input and output lines 1 ZKMA 2 ZKMB...
Page 23: ...12 INZ TN2ZKMA E 4 2 3 Allocation of the terminal block...
Page 24: ...13 INZ TN2ZKMA E 4 3 Wiring and piping diagram 4 3 1 1 Flow guide tube system...
Page 25: ...14 INZ TN2ZKMA E 4 3 1 2 Flow guide tube system with valve...
Page 26: ...15 INZ TN2ZKMA E 4 3 1 3 Ejector system...
Page 27: ...16 INZ TN2ZKMA E 4 3 1 4 Ejector system with valve...
Page 31: ...20 INZ TN2ZKMA E 5 2 Key operation flow diagram outline...
Page 32: ...21 INZ TN2ZKMA E...
Page 33: ...22 INZ TN2ZKMA E...
Page 34: ...23 INZ TN2ZKMA E...
Page 113: ...102 INZ TN2ZKMA E 11 SPECIFICATIONS 11 1 Specifications...
Page 114: ...103 INZ TN2ZKMA E...
Page 115: ...104 INZ TN2ZKMA E 11 2 Code symbols Converter...
Page 116: ...105 INZ TN2ZKMA E Detector...
Page 117: ...106 INZ TN2ZKMA E Ejector Exclusive cable...
Page 118: ......