6.4.2 Checking the error information
Display
message
State Remarks
Er1
Fault of heater temperature
Appears when control temperature of the heater exceeds
the set range.
The heater control is stopped.
Er2 Disconnection
detection
Appears when disconnection is detected at the sensor, or
thermocouples for temperature control.
The heater control is stopped.
Er3
Sensor error
Appears when the A/D value is saturated.
Er4
Span calibration error
Appears when the span calibration is abnormal.
(The calibration gas is unstable. / The calibration factor
setting is inappropriate.)
Er5
Zero calibration error
Appears when the zero calibration is abnormal.
(The calibration gas is unstable. / The calibration factor
setting is inappropriate.)
6.4.3 Checking the alarm information
Display
message
State Remarks
ALM
Oxygen concentration error
Appears when the oxygen concentration exceeds any of
specified HH / High / Lower / LL limit values.
(Refer to “10.5.4” to “10.5.8”)
H
High limit error
Appears together with ALM.
L
Lower limit error
Appears together with ALM.
HH
HH limit error
Appears together with ALM.
LL
LL limit error
Appears together with ALM.
You can select one of the following seven alarms to output to the alarm contact (Contact No. 21 and 22 of the
external terminal blocks) when an oxygen concentration error occurs.
(1)
[Not used]
: No alarm is output to the contact output.
(2)
[High limit alarm]
: Alarm contact is output when an high limit alarm occurs.
(3)
[Lower limit alarm]
: Alarm contact is output when a lower limit alarm occurs.
(4)
[HH limit alarm]
: Alarm contact is output when an HH limit alarm occurs.
(5)
[LL limit alarm]
: Alarm contact is output when a LL limit alarm occurs.
(6)
[High/lower limit alarm]
: Alarm contact is output when an high or lower limit alarm occurs.
(7)
[HH / LL limit alarm]
: Alarm contact is output when an HH or LL limit alarm occurs.
Summary of Contents for ZKMA
Page 1: ......
Page 13: ...2 INZ TN2ZKMA E 2 OPERATING PARTS AND THEIR FUNCTIONS 2 1 Outline Drawing 1 ZKMA 2 ZKMB...
Page 22: ...11 INZ TN2ZKMA E 4 2 2 Cable gland and input and output lines 1 ZKMA 2 ZKMB...
Page 23: ...12 INZ TN2ZKMA E 4 2 3 Allocation of the terminal block...
Page 24: ...13 INZ TN2ZKMA E 4 3 Wiring and piping diagram 4 3 1 1 Flow guide tube system...
Page 25: ...14 INZ TN2ZKMA E 4 3 1 2 Flow guide tube system with valve...
Page 26: ...15 INZ TN2ZKMA E 4 3 1 3 Ejector system...
Page 27: ...16 INZ TN2ZKMA E 4 3 1 4 Ejector system with valve...
Page 31: ...20 INZ TN2ZKMA E 5 2 Key operation flow diagram outline...
Page 32: ...21 INZ TN2ZKMA E...
Page 33: ...22 INZ TN2ZKMA E...
Page 34: ...23 INZ TN2ZKMA E...
Page 113: ...102 INZ TN2ZKMA E 11 SPECIFICATIONS 11 1 Specifications...
Page 114: ...103 INZ TN2ZKMA E...
Page 115: ...104 INZ TN2ZKMA E 11 2 Code symbols Converter...
Page 116: ...105 INZ TN2ZKMA E Detector...
Page 117: ...106 INZ TN2ZKMA E Ejector Exclusive cable...
Page 118: ......