6 - 28
INZ-TN2ZKJ-E
. Optical adjustment screen
For details of this item, refer to “Item 7.3.3
Optical zero adjustment method”.
The selenoid valve driving signal for zero
gas is turned ON by pressing the
ENT
key.
If the following operation is maladjusted, the measurement may be adversely and excessively
affected. Carry out the operation with utmost attention.
Caution
In values on the right side of screen, the moisture
interference for each component is already offset.
Description of moisture interference
adjustment screen
ENT
• Password Setting screen
It enables you to set a password to be used when
switching the parameter setting mode to the
maintenance mode.
Description of Password Setting screen
Note) The password set herein must be
managed for safety. Failure to enter
the correct password will not open
the Maintenance mode.
. Moisture interference adjustment screen
For details of this item, refer to “Item 7.3.4
Moisture interference adjustment method”.
By “
” or “
” key, move “ ” cursor to a
desired CH (component).
Make sure the gas for moisture interference
flows. By “
“ or “
” key, get the
value after moisture interference offset will
get near 0 and press the “
ENT
” key to
register the moisture interference corrected
value into the memory.