TN5FCXA2V5h-E
101
7.4
$GMXVWPHQWDIWHUUHSODFHPHQWRIWKHDPSOL¿HUXQLWRUWKHPHDVXULQJFHOO
Adjustment
After completion of the assembly work mentioned above, use the following procedures for
adjustment and setting. Adjustment should be performed using the HHC.
(1) After replacement of electronics unit (including replacement of internal parts)
No.
Item
Display symbol
RIORFDOFRQ¿JX
-
rator unit with
LCD display
(Relevant
page)
Display symbol of
HCC
(Relevant
page)
Contens of setting adjustment
1
TAG No.
1: TAG
(P44)
1: TAG No.
(P73)
Set the previous data before replacement of
amp unit.
2
Model code
2: TYPE
(P45)
2: TYPE
(P73)
Set the previous data before replacement of
amp unit.
3
Serial No.
3-1: SERIAL No.
3-2: VER
(P46)
3: SERIAL No.
(P74)
Not necessary for operation
4
Engineering unit
4: UNIT
(P47)
4: UNIT
(P74)
Set the previous data before replacement of
amp unit.
5
Range limit
5: URL
(P47)
5: RANGE LIMIT
(P75)
Not necessary for operation
6
Measuring range
6-1: LRV
6-2: URV
(P48)
6: RANGE
(P75)
Set the previous data before replacement of
amp unit.
7
Damping
7: DAMP
(P50)
7: DAMPING
(P76)
Set the previous data before replacement of
amp unit.
8
Output mode
8-1: OUT Md
8-2: CUT Pt
8-3: CUT Md
(P51)
8: OUTPUT MODE
(P77)
Set the previous data before replacement of
amp unit.
9
Burnout direction
9-1: BURNOUT
9-2: OVER
9-3: UNDER
(P53)
9: BURNOUT
(P78)
Set the previous data before replacement of
amp unit.
10
Zero/span
calibration
A-1: ZERO
A-2: SPAN
(P55)
A: CALIBRATE
(P879)
Implement span calibration after zero calibra-
tion.
11
Calibration output
circuit
b-1: 4mAAdj
b-2: 20mAAdj
b-3: FIXcur
(P57)
B: OUTPUT ADJ
(P80)
Loop check & calibrate fixed output current
(4mA,20mA)
12
M e a s u r e m e n t
data
(Normal mode)
C:DATA
(P81)
Check the measurement data.
13
Self-diagnosis
d1: AMPTMP
d2: ALMCHK
(P58)
D: SELF CHECK
(P81)
Check, if it is necessary.
14
Printer function
-
E: PRINT
(P82)
In case of HHC with printer option, print if it is
necessary.
15
L o c k o f a d j u s t -
ment functions
F: LOCK
(P59)
F: XMTR EXT.SW
(P82)
Set the previous data before replacement of
amp unit
16
Indication of
digital indicator
G-1: LDV
G-2: UDV
G-3: DP
G-4: LcdUnit
G-5: LcdOpt
(P60)
G: XMTR DISPLAY
(P83)
Set the previous data before replacement of
amp unit.
17
Programmable
linearization
function
-
H: LINEARIZE
(P86)
Set the previous data before replacement of
amp unit.
18
I n p u t - o u t p u t
range adjustment
I-1: LRVAdj
I-2: URVAdj
(P63)
I: RERANGE
(P88)
Adjust the input-output range (RERANGE) as
required.
19
Change of
saturation current
J-1: SAT LO
J-2: SAT HI
J-3: SPEC
(P65)
J: SATURE CUR
(P89)
Set the previous data before replacement of
amp unit.
20
Protective func-
tion of set value
K: GUARD
(P67)
K: WRITE PROTCT
(P90)
Set the previous data before replacement of
amp unit.
21
History informa-
tion
L-1: His ZERO
L-2: His SPAN
L-3: His CLEAR
L-4: His AMP
L-5: His CELL
(P68)
L: WRITE PROTCT
(P91)
Check data as necessary
(2) After replacement of detecting unit (including replacement of internal parts)
1
Zero/span
calibration
A-1: ZERO
A-2: SPAN
(P55)
A: CALIBRATE
(P79)
Implement span calibration after zero calibration