6.2 Commands and description
The following table lists the available commands and the associated passwords for read / write:
Command
Initial value
factory
setting
MODBUS
register-
address
Description
Password
Read
1
/ Write
2
10
40069
Averaging interval for wind speed
and wind direction.
None
User
96
40005
Select baud rate.
None
User
0
40081
Mode of calculation for twilight.
None
User
0
40037
North correction of wind direction.
None
User
1
40001
Quick-start mode.
None
User
5
40035
Heating capacity anti-condensation
None
User
0
(INTERNER
BUS)
1 (MODBUS)
40003
Identification number resp. slave
address.
None
User
0
40009
Set key / password.
None
None
0
40045
LED control.
None
None
-
40029
Reset.
None
User
0
40071
Station height.
None
User
-
45005
SW-Version.
None
-
-
-
Telegram output.
None
None
0
-
Automatic telegram output.
None
None
0
40073
Time zone.
None
User
Table 4 : List of commands
1
: Command without parameter (used to read selected parameter).
2
: Command with parameter (used to write a new parameter).
6.2.1 Command AI
<id>AI<parameter><CR>
Averaging interval for wind speed and wind direction
Access:
Read / write.
Description:
The command AI is used to specify the averaging interval for the wind
speed and wind direction in minutes.
The averaging of the wind velocity is scalar, and the averaging of the
wind direction vectorial.
If the parameter is 0, the averaging is deactivated, and the mean
values correspond to the instantaneous values. Here, the wind
direction is reset (to 0°) during calm (< 0.6m/s).
Wind from the North is displayed with 360°.
Parameter description:
AI = 0
averaging disabled
AI = 1
averaging interval = 1 minute
Value range:
0...10