3. Guidance and manufacturer’s declaration - electromagnetic immunity
This medical LCD monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of monitor should assure that it is used in such an environment.
Immunity test
IEC 60601 Test
level
Compliance level Electromagnetic environment-guidance
Power frequency
(50/60Hz)
Magnetic field
IEC 61000-4-8
3.0 A/m
3.0 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
Voltage dips,
short Interrup-
tions and voltage
variations on
power supply
input lines
IEC 61000-4-l 1
<5%
U
T
(>95% dip in
U
T
) for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycle
70%
U
T
(30% dip in
U
T
)
for 25 cycle
<5%
U
T
(<95% dip in
U
T
) for 5 sec.
<5%
U
T
(>95% dip in
U
T
) for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycle
70%
U
T
(30% dip in
U
T
)
for 25 cycle
<5%
U
T
(<95% dip in
U
T
) for 5 sec.
Main power quality should be that of a
typical commercial or hospital environment.
If the user of monitor requires continued
operation during power mains interruptions,
it is recommended that monitor be powered
from an uninterruptible power supply or a
battery.
Note:
U
T
is the A.C. mains voltage prior to
application of the test level.
Radiated RF
IEC 61000-4-3
3 V/m
80.0 MHz to 2.5
GHz
3 V/m
80.0 MHz to 2.5
GHz
Recommended separation distance
[
] √
80MHz to
800MHz
[
] √
80MHz to
2.5GHz
Where
P
is the maximum output power rat-
ing of the transmitter in watts (W) according
to the transmitter manufacturer and
d
is the
recommended separation distance in meters
(m). Field strengths from fixed RF transmit-
ters, as determined by an electromagnetic
site survey, should be less than the compli-
ance level in each frequency range.