FSN1962 - 6/2015
English - 11
1. Guidance and manufacturer’s declaration - electromagnetic emission
The medical LCD monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the medical LCD monitor should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment -guidance
RF Emissions
CISPR 11
Group 1
The medical LCD monitor uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The medical LCD monitor is suitable for use in all establish-
ments, including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
D
Voltage fluctuations
IEC 61000-3-3
Complies
2. Guidance and manufacturer’s declaration - electromagnetic immunity
This medical LCD monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the medical LCD monitor should assure that it is used in such an environment.
Immunity test
IEC 60601 Test
level
Compliance
level
Electromagnetic environment-guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity should
be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/
output lines
2 kV for power
supply lines
1 kV for input/
output lines
Main power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential mode
2 kV common mode
Main power quality should be that of a typical
commercial or hospital environment.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80MHz
3 Vrms
150 kHz to
80MHz
Portable and mobile RF communications equip-
ment should be used no closer to any part of
the LCD monitor, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance : d
[
] √
Where
P
is the maximum output power rating of
the transmitter in watts (W)