Installation and start-up
3.1
Scope of delivery
The scope of delivery includes:
>
1 x DS01+ pressure transducer
>
1 x quick reference guide
>
optional: 1 x calibration certificate
3.2 Assembly
The DS01+ pressure transmitter can be attached to the wall via the side lugs using dowels and screws.
Notice
>
Mount the device on a vertical, smooth surface
>
Do not mount the device in the immediate vicinity of sources of interference or heat
3.3 Start-up
3.3.1 Connection
Connect the supply voltage as well as the output wiring according to the connection diagram. For this purpose run the cables
through the cable glands attached to the housing and connect them to the screw terminals provided.
In order to guarantee the specified accuracy, the sensor should only be fully put into operation after a run-in time of one
hour. After the run-in time has elapsed, a zero point correction should be carried out to adjust the pressure transmitter to the
ambient pressure (
see 4.1 Zero point correction
).
3.3.2 Setting the measuring range
+
-
Clamps
0V
V+
I
U
0V
DIP-Switches
S1
S2
S3
S4
S5
S6
Button S
On
Off