Configuration Menu
Top sensor offset (unused on those models)
Bottom sensor offset
Permits to change the temperature displayed by another
measure. Let's get the value actually measured in stabilized
conditions. Do not enter a temperature difference.
Door open alarm delay (e.g. 00h00min30s).
Low temperature alarm threshold. Fix the chosen value of low
temperature threshold.
High temperature alarm threshold. Fix the chosen value of high
temperature threshold.
Back-up injection configuration: OFF / CO
2
/ N
2
.
More informations below.
Back-up injection trigger threshold (e.g. -60°C).
More informations below.
Start-up delay time (e.g. 00h00min30s).
More informations below.
Delay door open alarm to dry contact (e.g. 00h00min30s).
More informations below.
Return to the main display
Co
2
/ N
2
injection threshold setting (option)
To allow back-up of the injection, CO2 or N2 injection must be selected and injection threshold must be set.
The CO
2
/ N
2
injection is inactive until the freezer temperature has stabilized after installation.
Back-up injection configuration: OFF / CO
2
/ N
2
Select the corresponding mode.
Emergency injection trigger threshold
Set the desired temperature.
When the freezer is installed, injection must be enabled only once the temperature set point has been
reached, otherwise the gas cylinders will be discharged during the first temperature drop.
Start-up delay setting