35
Using the device
5.2.3
Adjusting the funnel height
Depending on the flowing properties and size of the material,
a smaller gap (for fine powder with good flowing properties)
or a larger gap (for coarse material) between the funnel and
the feeder is needed. The gap to the feeder should be about
three times the maximum diameter of the material.
Turn the knurled screw on the top of the funnel support column in order to
move the funnel up and down.
5.3
Wet measurement
5.3.1
Inserting the measuring cell
1.
Open the housing flap.
2.
Unlock the centring plate and hinge it backwards.
3.
Close the housing flap again.
4.
Place the wet dispersion unit in front of or next to the ANALYSETTE 28
ImageSizer.
5.
Insert the wet measuring cell in the ANALYSETTE 28 ImageSizer .