Installation
4.6
Electrical connection
DANGER
Provide short-circuit protection!
Risk of damage due to short-circuits.
−
Make sure that the socket is connected to a mains line
protected with a residual current circuit breaker.
DANGER
Mains voltage!
Changes to the connection line may only be made by a quali-
fied person.
Before establishing the connection, compare the voltage and current values
stated on the type plate with the values of the power supply system to be used.
CAUTION
Ignoring the values on the type plate may result in damage to
the electrical and mechanical components.
CAUTION
Ensure that the mains supply provides a functional device
ground connection!
4.6.1
Stability of the mains voltage
Instruments with electronic components require a stable supply voltage (+/-
10% deviation). For weak or non-interference free mains supplies (voltage peaks
through inductive load changes or switched-mode power supplies), we recom-
mend connecting via a voltage stabiliser and filter (Art. No. 20.6000.00).
4.7
Preparation of the ANALYSETTE 28 ImageSizer
Open the front flap (7) forwards by lifting up both latches (6).