![Fresenius Medical Care 460018 Operator'S Manual Download Page 40](http://html.mh-extra.com/html/fresenius-medical-care/460018/460018_operators-manual_2331648040.webp)
34
P/N 460018 Rev. D
Unit Name & Location #
Operator
Date
Time
Dry Acid Dissolution Unit Serial #
Batch #
Do ALL of the following
(check when completed): Empty Tank Rinse Tank Verify Expirations
FLUSH
the transfer hose with 3.5 Gallons.
FILL
a graduated cylinder from hose.
MEASURE
temperature of liquid in cylinder to one decimal place. Enter value here:
MEASURE
the Specific Gravity value*. Enter value here:
CIRCLE BOTH
temperature range for the dry acid used
AND
the
CHECK
one of the
Low / High Specific gravity values for the catalog number below:
following statements
regarding the measured
Specific gravity:
FAILED HIGH.
Value was
Above circled range.
PASSED.
Value was within
Circled table range.
FAILED LOW.
Value was
Below circled table range.
Operator:
Reviewer:
Signature
Signature
Minimum Thermometer Specifications: Temperature Range 25°C +/- 10°C (59° to 95° F) and ac/- 1°C (3.6 °F)
*Only use a hydrometer calibrated at 60
0
F following ASTM Standard E126.
Form 2: Citrasate
®
Dry – 132 Gallon Batch Production Record
ADD
Dry Acid Products and
RECORD
case information below:
Dry Acid Catalog #
BOX Lot #
if 4
th
bag used
CASE 1:
CASE 2:
CASE 3:
CASE 4:
CASE 5:
CASE 6:
CASE 7:
CASE 8:
** After Final Fill Level has been reached,
CLOSE
water supply valve to the Unit then
CHECK HERE:
16.5°C to 21.4°C
(61.7°F to 70.6°F)
21.5°C to 26.4°C
(70.7°F to 79.6°F)
26.5°C to 31.4°C
(79.7°F to 88.5°F)
Catalog Number
Low
High
Low
High
Low
High
1K
0FD1251-3B
1.191
1.203
1.188 1.200
1.186
1.198
2K
0FD2201-3B
1.191
1.203
1.189 1.201
1.187
1.199
0FD2231-3B
1.192
1.204
1.190 1.202
1.187
1.199
0FD2251-3B
1.192
1.204
1.190 1.202
1.188
1.200
0FD2301-3B
1.193
1.205
1.191 1.203
1.189
1.201
3K
0FD3201-3B
1.193
1.205
1.191 1.203
1.188
1.200
0FD3231-3B
1.194
1.206
1.191 1.203
1.189
1.201
0FD3251-3B
1.194
1.206
1.192 1.204
1.189
1.201
0FD3301-3B
1.195
1.207
1.193 1.205
1.190
1.202
1
2
3
4
5
6
7
8
0
C
Summary of Contents for 460018
Page 1: ... DRY ACID DISSOLUTION UNIT 132 GALLONS 500 LITERS OPERATOR S MANUAL P N 460018 REV D ...
Page 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 37: ...P N 460018 Rev D 31 PAGE INTENTIONALLY LEFT BLANK ...
Page 39: ...P N 460018 Rev D 33 PAGE INTENTIONALLY LEFT BLANK ...
Page 41: ...P N 460018 Rev D 35 PAGE INTENTIONALLY LEFT BLANK ...
Page 43: ...P N 460018 Rev D 37 PAGE INTENTIONALLY LEFT BLANK ...
Page 45: ...P N 460018 Rev D 39 PAGE INTENTIONALLY LEFT BLANK ...
Page 50: ...44 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 51: ...P N 460018 Rev D 45 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Page 52: ...46 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 53: ...P N 460018 Rev D 47 ...
Page 54: ...48 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 55: ...P N 460018 Rev D 49 ...
Page 56: ...50 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 57: ...P N 460018 Rev D 51 ...
Page 58: ...52 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 59: ...P N 460018 Rev D 53 ...
Page 60: ...54 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 61: ...P N 460018 Rev D 55 ...
Page 62: ...56 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 63: ...P N 460018 Rev D 57 ...
Page 64: ...58 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 65: ...P N 460018 Rev D 59 ...
Page 66: ...60 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 68: ...62 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 70: ...64 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 75: ...P N 460018 Rev D 69 PAGE INTENTIONALLY LEFT BLANK ...