7 . 1 S P E C I F I C G R A V I T Y T E S T
18
P/N 460018 Rev. D
Filling Hydrometer Cylinder slowly down the
side of cylinder
4.
Slightly OPEN the Main Transfer Ball Valve
next to the Filter Housing (Figure 12, item A).
5.
Flush the Transfer Hose:
WARNING!
F
AILURE TO FLUSH THE TRANSFER HOSE WITH
3.5
GALLONS PRIOR TO MEASURING SPECIFIC GRAVITY MAY
RESULT IN AN INCORRECT SPECIFIC GRAVITY READING
,
AND
/
OR SUPPLY OF AN INCORRECT CONCENTRATE
FORMULATION TO THE
H
EMODIALYSIS
M
ACHINE
,
WHICH
MAY RESULT IN PATIENT INJURY OR DEATH
.
WARNING!
T
HE FLUSH SOLUTION IS A MIXTURE OF
CONCENTRATE FROM THE PREVIOUS BATCH AND THE NEW
BATCH
.
T
HE COMPOSITION OF THE FLUSH SOLUTION MUST
BE CONSIDERED WHEN DETERMINING CLINIC POLICIES FOR
DISPOSITION OF THE FLUSH SOLUTION
.
S
UPPLY OF AN
INCORRECT CONCENTRATE FORMULATION TO THE
H
EMODIALYSIS
M
ACHINE MAY RESULT IN PATIENT INJURY
OR DEATH
.
•
Press Dissolution START button, slightly open Transfer Nozzle Valve and
flush approximately 3.5 gallons.
•
Once the 3.5 gallons of solution have been flushed out of the transfer hose,
CLOSE the Transfer Nozzle.
•
Disposition the flushed solution according to clinic policies and procedures.
6.
Fill the Hydrometer Cylinder:
•
Insert the Transfer Nozzle into the Hydrometer Cylinder.
•
Slowly OPEN the Transfer Nozzle Valve until solution starts to
slowly accumulate into the Hydrometer Cylinder. Allowing
solution to fill down the side of the beaker minimizes the
creation of bubbles within the solution (Figure 12, item B).
•
Fill the Hydrometer Cylinder approximately 2/3 rd full.
•
CLOSE Transfer Nozzle and then press PAUSE on the Display
Panel.
•
Place Transfer Nozzle onto a clean surface.
7.
Make sure the solution you gathered does not have excessive
amount of bubbles. To release excess amount of bubbles from
the solution gently tap the Hydrometer Cylinder or gently swirl
the entire Hydrometer Cylinder until bubbles have been
released from the solution (Figure 13).
B
A
Figure 12
Figure 13
Summary of Contents for 460018
Page 1: ... DRY ACID DISSOLUTION UNIT 132 GALLONS 500 LITERS OPERATOR S MANUAL P N 460018 REV D ...
Page 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 37: ...P N 460018 Rev D 31 PAGE INTENTIONALLY LEFT BLANK ...
Page 39: ...P N 460018 Rev D 33 PAGE INTENTIONALLY LEFT BLANK ...
Page 41: ...P N 460018 Rev D 35 PAGE INTENTIONALLY LEFT BLANK ...
Page 43: ...P N 460018 Rev D 37 PAGE INTENTIONALLY LEFT BLANK ...
Page 45: ...P N 460018 Rev D 39 PAGE INTENTIONALLY LEFT BLANK ...
Page 50: ...44 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 51: ...P N 460018 Rev D 45 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Page 52: ...46 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 53: ...P N 460018 Rev D 47 ...
Page 54: ...48 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 55: ...P N 460018 Rev D 49 ...
Page 56: ...50 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 57: ...P N 460018 Rev D 51 ...
Page 58: ...52 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 59: ...P N 460018 Rev D 53 ...
Page 60: ...54 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 61: ...P N 460018 Rev D 55 ...
Page 62: ...56 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 63: ...P N 460018 Rev D 57 ...
Page 64: ...58 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 65: ...P N 460018 Rev D 59 ...
Page 66: ...60 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 68: ...62 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 70: ...64 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 75: ...P N 460018 Rev D 69 PAGE INTENTIONALLY LEFT BLANK ...