P/N 460018 Rev. D
5
S
ECTION
4:
W
ARNINGS AND
C
AUTIONS
WARNING
!
N
EVER PUT YOUR HEAD INTO THE
D
RY
A
CID
D
ISSOLUTION
U
NIT
T
ANK
.
T
HIS MAY RESULT IN OPERATOR
FALLING INTO TANK
.
WARNING
!
I
N THE EVENT OF A FOREIGN OBJECT FALLING INTO THE TANK DURING ANY ACTIVE OPERATION
,
THE
OPERATION SHOULD BE
PAUSED.
I
F NEEDED
,
USE
STEP
MODE
TO PLACE
U
NIT INTO A
D
RAIN
O
PERATION TO EMPTY
THE TANK
.
I
F CONCENTRATE SOLUTION NEEDS TO BE DISPOSED
,
MAKE SURE TO COMPLY WITH LOCAL
,
STATE AND FEDERAL
REQUIREMENTS
.
A
FTER DRAIN
O
PERATION
,
UNPLUG
D
RY
A
CID
D
ISSOLUTION
U
NIT FROM ITS POWER SOURCE
.
A
REACH
TOOL IS RECOMMENDED FOR REMOVING FOREIGN OBJECT OUT OF TANK
.
H
OWEVER
,
SHOULD THE FOREIGN OBJECT BE
UNATTAINABLE
,
THEN
S
EE
S
ECTION
13:
D
ISPOSAL OF
C
ONCENTRATE
S
OLUTION
,
PAGE
WARNING
!
I
F THE DRY ACID DISSOLUTION UNIT IS RELOCATED
,
THEN A SAMPLE MUST BE DRAWN FROM THE FIRST BATCH
OF CONCENTRATE MADE AND IT MUST BE ANALYZED FOR CORRECT SOLUTION MIX BEFORE THE CONCENTRATE CAN BE
USED
.
P
RIOR TO RELOCATION
,
REQUEST TWO
(2)
EMPTY SAMPLE BOTTLES
(P/N
G83-535-02)
THEN SEE
A
PPENDIX
D:
F
IRST BATCH VERIFICATION FOR FURTHER INSTRUCTIONS ON WHEN AND WHERE TO SEND SAMPLE OF THE FIRST BATCH OF
CONCENTRATE
.
WARNING!
T
HE USE OF EYE PROTECTION AND GLOVES IS RECOMMENDED WHEN HANDLING DRY ACID PRODUCT
.
I
F
CONTACT WITH EYES
,
RINSE IMMEDIATELY FOR
15
MINUTES
.
I
F CONTACT WITH SKIN
,
FLUSH WITH PLENTY OF SOAP AND
WATER
.
S
EE
M
ATERIAL
S
AFETY
D
ATA
S
HEETS
(MSDS)
FOR THE
D
RY
A
CID
P
RODUCT BEING USED FOR FURTHER
P
ERSONAL
P
ROTECTIVE
E
QUIPMENT
(
PPE
)
OR
E
MERGENCY REQUIREMENTS
/
INSTRUCTIONS
.
WARNING!
S
HOULD THE
D
RY
A
CID
D
ISSOLUTION
U
NIT
F
AIL TO COMPLETE ANY OPERATION WITHIN ANY
C
YCLE DO NOT
PROCEED THROUGH CYCLE USING
S
TEP
M
ODE
.
C
ALL
T
ECHNICAL
S
ERVICE FOR FURTHER ASSISTANCE
(1-800-227-2572).
WARNING!
F
AILURE TO INSTALL
,
OPERATE
,
AND MAINTAIN THIS EQUIPMENT ACCORDING TO THE MANUFACTURER
’
S
INSTRUCTIONS MAY CAUSE PATIENT DEATH OR INJURY
.
WARNING!
T
HIS
D
RY
A
CID
D
ISSOLUTION
U
NIT MAY BE SUSCEPTIBLE TO ELECTROMAGNETIC INTERFERENCE
(EMI).
D
EVICES EMITTING ELECTROMAGNETIC RADIATION SUCH AS ANALOG PORTABLE PHONES
,
RADIO EQUIPMENT
(
WALKIE
-
TALKIES
,
ETC
.),
RADIO TRANSMITTERS
,
AND LIKE EQUIPMENT
,
SHOULD NOT BE USED IN THE VICINITY OF THIS EQUIPMENT
.
T
HE OPERATOR SHOULD MONITOR THE FUNCTION OF THE MACHINE AND REMOVE UNNECESSARY EQUIPMENT FROM THE
TREATMENT AREA SHOULD THESE EVENTS OCCUR
.
WARNING!
T
HE ELECTRICAL SOURCE MUST BE SINGLE PHASE
,
THREE
-
CONDUCTOR TYPE PROVIDED WITH A GROUND
FAULT INTERRUPTER AT
120V
AC
,
60
H
Z
.
T
HE PROPER POLARITY AND GROUND INTEGRITY MUST BE INITIALLY CHECKED
AND MAINTAINED
.
F
AILURE TO DO SO MAY RESULT IN ELECTRICAL SHOCK OR BURN TO THE OPERATOR
.
WARNING!
S
HOCK HAZARD
.
D
O NOT REMOVE COVERS
.
R
EFER SERVICING TO QUALIFIED PERSONNEL
.
R
EPLACE FUSES
ONLY WITH THE SAME TYPE AND RATING
.
WARNING!
F
AILURE TO FLUSH THE TRANSFER HOSE WITH
3.5
GALLONS PRIOR TO MEASURING SPECIFIC GRAVITY MAY
RESULT IN AN INCORRECT SPECIFIC GRAVITY READING
,
AND
/
OR SUPPLY OF AN INCORRECT CONCENTRATE FORMULATION TO
THE
H
EMODIALYSIS
M
ACHINE
,
WHICH MAY RESULT IN PATIENT INJURY OR DEATH
.
S
EE SECTION
7.1,
#5,
PAGE
WARNING!
F
OLLOWING DISINFECTION
,
TWO
(2)
COMPLETE
RINSE
CYCLES
MUST BE PERFORMED AND THE UNIT
TESTED TO ENSURE THE ABSENCE OF RESIDUAL BLEACH
(S
EE
S
ECTION
8.3 Sodium Hypochlorite (Regular Bleach)
Disinfection, page 23).
WARNING!
F
OLLOW THE
M
ANUFACTURER
’
S INSTRUCTIONS WHEN MEASURING THE
SPECIFIC GRAVITY
OF A SAMPLE OF
THE MIXED
F
RESENIUS
M
EDICAL
C
ARE
D
RY
A
CID
P
RODUCT
(S
EE
F
ORM
1
FOR
G
RANU
F
LO
®
P
RODUCT
,
OR
F
ORM
2
FOR
C
ITRASATE
DRY
P
RODUCT
,
WITH A HYDROMETER CALIBRATED AT
60°F
FOLLOWING
ASTM
S
TANDARD
E126
PRIOR TO TRANSFERRING TO A DIALYSIS MACHINE OR HOLDING TANK
.
WARNING!
C
ONNECT WATER INLET ACCORDING TO THE SPECIFICATIONS FOR THE
D
RY
A
CID
D
ISSOLUTION
U
NIT
.
T
HE
WATER USED MUST MEET
ANSI/AAMI
OR
ISO
STANDARDS FOR WATER USED IN HEMODIALYSIS
(
CURRENTLY
ANSI/AAMI
RD62
,
OR
ISO 13959
).
T
HE CORRECT IONIC CONCENTRATION AND BACTERIAL QUALITY CAN GENERALLY BE ACHIEVED IN
THE DIALYSATE ONLY WITH TREATED WATER
.
B
E SURE THAT ALL SPECIFICATIONS ARE SATISFIED
.
T
HE WATER SOURCE
MUST BE MONITORED PERIODICALLY TO DETECT FLUCTUATIONS IN WATER COMPOSITION AND QUALITY THAT COULD HAVE
AN ADVERSE EFFECT ON THE PATIENT
,
HEMODIALYSIS MACHINE OR
D
RY
A
CID
D
ISSOLUTION
U
NIT
.
P
ARTICULAR ATTENTION
MUST BE TAKEN FOR CHEMICALS SUCH AS ALUMINUM
,
CHLORINE
,
AND CHLORAMINES
,
AS THESE CHEMICALS CAN CAUSE
Summary of Contents for 460018
Page 1: ... DRY ACID DISSOLUTION UNIT 132 GALLONS 500 LITERS OPERATOR S MANUAL P N 460018 REV D ...
Page 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 37: ...P N 460018 Rev D 31 PAGE INTENTIONALLY LEFT BLANK ...
Page 39: ...P N 460018 Rev D 33 PAGE INTENTIONALLY LEFT BLANK ...
Page 41: ...P N 460018 Rev D 35 PAGE INTENTIONALLY LEFT BLANK ...
Page 43: ...P N 460018 Rev D 37 PAGE INTENTIONALLY LEFT BLANK ...
Page 45: ...P N 460018 Rev D 39 PAGE INTENTIONALLY LEFT BLANK ...
Page 50: ...44 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 51: ...P N 460018 Rev D 45 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Page 52: ...46 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 53: ...P N 460018 Rev D 47 ...
Page 54: ...48 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 55: ...P N 460018 Rev D 49 ...
Page 56: ...50 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 57: ...P N 460018 Rev D 51 ...
Page 58: ...52 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 59: ...P N 460018 Rev D 53 ...
Page 60: ...54 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 61: ...P N 460018 Rev D 55 ...
Page 62: ...56 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 63: ...P N 460018 Rev D 57 ...
Page 64: ...58 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 65: ...P N 460018 Rev D 59 ...
Page 66: ...60 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 68: ...62 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 70: ...64 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 75: ...P N 460018 Rev D 69 PAGE INTENTIONALLY LEFT BLANK ...