![Fresenius Medical Care 460018 Operator'S Manual Download Page 33](http://html.mh-extra.com/html/fresenius-medical-care/460018/460018_operators-manual_2331648033.webp)
P/N 460018 Rev. D
27
S
ECTION
9:
M
ANUAL
C
ONTROL
O
PERATIONS
9.1 P
AUSE
S
TATE
The PAUSE State function is to allow for special occurrences in the operation of the Dry Acid Dissolution
Unit. For example; if the unit is not working properly or the operator needs to PAUSE a timed operation,
then the PAUSE button may be pressed.
This will cause the Indicating Light for the current step of the
operation to flash. Pressing the PAUSE button will disable the
pump, drain valve, fill valve, etc. To continue the cycle, press
the START button and the timers will resume.
The FILL Operation cannot be paused.
CAUTION!
D
O NOT ALLOW THE
U
NIT TO REMAIN FULL OF WATER WITHOUT THE ADDITION OF
D
RY
A
CID
P
RODUCT
.
BACTERIAL GROWTH MAY OCCUR
.
9.2
S
TEP
M
ODE
CAUTION!
STEP
MODE
FUNCTION IS
NOT
TO BE USED TO SPEED UP THE DISSOLUTION OR RINSE
O
PERATIONS
.
The STEP MODE function is intended to be used during the Disinfection Operation or when it is necessary
to discard an incorrectly mixed batch of solution. The STEP MODE function has the ability to skip critical
steps in the Dissolution and Rinse Cycles but using this mode during these cycles can cause unintended
operation of the Dry Acid Dissolution Unit resulting in an incorrectly mixed solution.
To enter the STEP MODE press the STEP MODE ON/OFF
button until the Step Mode Indicator Light illuminates. The
system will enter the STEP MODE and all operations will be
suspended.
Press the STEP button to the desired operation. When the
desired operation light becomes illuminated, press the
ON/OFF button and the operation is continued.
CAUTION!
W
HEN THE
STEP
Light
IS
OFF,
THE
D
RY
A
CID
D
ISSOLUTION
U
NIT WILL RESUME OPERATION
.
Summary of Contents for 460018
Page 1: ... DRY ACID DISSOLUTION UNIT 132 GALLONS 500 LITERS OPERATOR S MANUAL P N 460018 REV D ...
Page 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 37: ...P N 460018 Rev D 31 PAGE INTENTIONALLY LEFT BLANK ...
Page 39: ...P N 460018 Rev D 33 PAGE INTENTIONALLY LEFT BLANK ...
Page 41: ...P N 460018 Rev D 35 PAGE INTENTIONALLY LEFT BLANK ...
Page 43: ...P N 460018 Rev D 37 PAGE INTENTIONALLY LEFT BLANK ...
Page 45: ...P N 460018 Rev D 39 PAGE INTENTIONALLY LEFT BLANK ...
Page 50: ...44 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 51: ...P N 460018 Rev D 45 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Page 52: ...46 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 53: ...P N 460018 Rev D 47 ...
Page 54: ...48 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 55: ...P N 460018 Rev D 49 ...
Page 56: ...50 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 57: ...P N 460018 Rev D 51 ...
Page 58: ...52 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 59: ...P N 460018 Rev D 53 ...
Page 60: ...54 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 61: ...P N 460018 Rev D 55 ...
Page 62: ...56 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 63: ...P N 460018 Rev D 57 ...
Page 64: ...58 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 65: ...P N 460018 Rev D 59 ...
Page 66: ...60 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 68: ...62 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 70: ...64 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Page 75: ...P N 460018 Rev D 69 PAGE INTENTIONALLY LEFT BLANK ...