Technical information
7.1
Performance
Weight
Pump:
550 g
Pump holder:
500 g
Dimensions
Pump: 132 x 120 x 45 mm
Holder: 146 x 162 x 115 mm
Disposal
For appropriate disposal, at the end of the
device life, please contact your local
Fresenius Kabi
organisation or your local
distributor.The life cycle of the pump is 5
years provided that the maintenance is
properly performed as described on Section
9.2, page 51.
Protection against electric shock
Protection class II
Type CF applied part
- the
activ
set
- the
Ambix
activ
pump
- the
activ
Holder
Electromagnetic interference
This device can be disturbed by large
electromagnetic fields, external electrical
influences and electrostatic discharges above
the limits stipulated by EN 60601-1-2 and EN
60601-2-24. If you wish to use the device in
special conditions, please contact your local
Fresenius Kabi branch.
This mobile RF communication equipment
can affect medical electrical equipment.
Safety of ElectroMedicalEquipment
Conforms to EN/IEC 60601-1 &
EN/IEC 60601-2-24
Protection against moisture
Pump IP34 and Holder IP33 (splash-
protected)
Electrical supply
Power supply:
100-230 V ±10% / 50-60 Hz 14 VA
Pump holder output: 7.75 V/800 mA
Battery type: NiMH 4.8 V 1.8 Ah
(Nickel-Metal Hydride)
Battery life: 40 hours at 125 ml/h
The maximum battery life could only be
achieved after several charge/discharge
cycles.
Avoid short circuit and excessive
temperature.
Equipment group
Class IIb
(Medical Devices Directive 93/42/EEC)
Class III
(Canadian Medical Devices Regulations
SOR/98-282)
Environmental conditions of transport and
storage between uses:
Temperature:-20°C to 45°C
Relative humidity: 10 to 85%, no
condensation
Atmospheric pressure: 700 to 1060 hPa
Pump, pump holder operating conditions:
Temperature: +13 to +40°C
Relative humidity: 20 to 85%, no
condensation
Atmospheric pressure: 700 to 1060 hPa
Altitude: max 3000m
Attachment size (vertical/horizontal)
cylinder shape: 18-36 mm
square shape: 10x25 mm
Deviation of rate
± 5% at 22.5°C ± 2.5°C, pressure < 0.1 bar,
flow rates above 10 ml/h
Downstream occlusion detection
Occlusion pressure (downstream), 2 levels:
L. approx. 0.5 bar
H. approx. 0.8 bar
KVO:
0 - 10 ml/h
preset 2 ml/h
Infusion program:
Continuous, volume/time, ramp, intermittent
Nurse call
Potential-free switch, 4 KV, decoupling
Power output: 24 V / 100 mA to power Nurse
Call
*
Serial communication
RS232 optical isolation 4 kV
*
Communication format: 9600 baud / Even
parity / 7 data bits / 1 stop bit
*
Connected only to SELV circuits (<60VDC)