29
3.6.2
Enter the parameter setting menu
3.6.3
Password for the parameter setting menu
3.6.4
KVO rate
Information
Pressing either the i/OK or STOP/ESC button returns the user to the
parameter menu and the number of the function that has been selected
flashes in the display.
Press the Lock button ( ) and the Alarm silence button (
)
simultaneously to enter the parameter menu.
Before changing any parameter, the password has to be entered by
selecting option 1 from the menu. If the password is not entered, it is only
possible to review the parameter values. The preset of the password is
3. The password can be altered, see Section 3.6, page 28.
Button
Action
Upper +/–
Select 1 from the menu to enter the password (see list
Lower +/–
Enter password (can be set between 0 and 250, preset
is 3)
i/OK
Confirm the password
The
Ambix
activ
can perform a Keep Vein Open (KVO) function. This
function will continue the infusion at a low rate after the target volume is
reached in order to keep the vascular access device and vein open. The
KVO function is available in all modes except continuous mode with
deactivated target volume. The setting of the KVO is between 0
(deactivated) and 10 ml/h in increments of 1 ml/h. The preset KVO rate
is 2 ml/h. The KVO rate will not exceed the set flow rate for the infusion,
irrespective of the configured default value for KVO.
When the KVO function is activated and the target volume selected,
KVO will be displayed in the display. During KVO application, KVO will
flash in the display and the KVO flow rate will be displayed in the upper
row.
Information
In case of KVO interruption, the function will be deactivated.
In case of KVO change, the parameters values are reset to the
factory initial settings.