3.4.11
Keypad lock
3.4.12
Pump reset
3.4.13
Target volume
Lock the keypad of the pump prevents unintentional change of infusion
settings of the pump.
Press the Lock button once. A key symbol will flash in the display for 5
seconds. If the OK button is pressed during this time, the keypad will be
locked and the key symbol will remain in the display. The locked keypad
will be indicated by a beep if one of the locked buttons is pressed.
To unlock the keypad press the Lock button and then within 5 seconds
press the OK button.
Information
Once the keypad lock is activated, only the following buttons are active:
On/Off button:
Switches pump on and off.
Start button:
Continues the infusion after an alarm without unlocking the
keypad.
Stop/Esc button:
Stops administration when pump is running (an alarm will appear).
Stop/Esc button:
Clears the alarm permanently.
Alarm silence button:
Silences alarms for two minutes.
Info button:
Accesses info function. To exit the info function, wait for 5
seconds.
Press the upper + and – buttons simultaneously for more than 3 seconds
to reset the pump and set all values to the factory default settings.
In continuous mode, the target volume feature can be
activated/deactivated by pressing the lower + and – buttons
simultaneously.
Please make sure to program a target volume 5% lower than the volume
in the container.