3.6.9
Interval time of the alarm
Information
The volume for the pre-alarm is silent without any further
reminder by pressing the Alarm silence button.
In Volume over Time, Intermittent and Ramp modes, the target
volume feature is automatically switched on.
The alarm will sound 30 minutes before the end of the target
volume even if the programmed percentage threshold is not
reached.
The alarm is deactivated when the value is 0.
Ensure that the setting is appropriate for the patient.
Button
Action
Upper +/–
Select 8 from the menu to enable change to
percentage of volume for end alarm
Lower +/–
Change to desired value (0 - 100%). Preset is 10%.
i/OK
Confirm the adjusted value
The time between alarms can be adjusted to between 5 and 300
seconds in increments of 5 seconds.
Button
Action
Upper +/–
Select 15 from the menu to enable change to alarm
interval time
Lower +/–
Change to desired value (between 5 and 300 s). Preset
is 30s.
i/OK
Confirm the adjusted value