background image

 

43 

 

 
Section 7.8 Settings Submenu Screen 

 

7.81Settings Screen 

 
The settings screen allows you to adjust pump motor speed (on vinegar buffered systems only) as 
well as the screen brightness and audio volume. This screen also contains the serial number, the 
software version number, and display version number. These numbers are required for customer 
service and technical support. 

 

7.82 Screen Settings 

 

The screen brightness and audio volume can be adjusted to suit your needs. From the Settings 
submenu touch the “Screen Settings” button to display the adjustment screen. 

 
Simply touch the Plus or Minus buttons to adjust as needed. Touch the “Go Back” button to return to 
the settings screen, or Home to return to the Home / Main run screen. 
 
 
 

 
 
 
 

 

 

Summary of Contents for FX-CL-F

Page 1: ...Document No OMFXCLFT042018 1 2018 Foxcroft Equipment Service Co Inc Model FX CL F T Reagentless Amperometric Free or Total Chlorine Residual Analyzer Instruction Manual...

Page 2: ...2...

Page 3: ...lectrode Finger Tip 22 2 5 Sensor Maintenance Storage 23 Section 3 System Component Identification Description 3 0 Model FX CL F T Specifications 24 3 1 Analyzer Component Identification Description 2...

Page 4: ...Screen 43 7 82 Screen Settings 43 Section 8 Troubleshooting 44 8 1 Specific Troubleshooting Of The Sensor 45 8 11Testing the leak tightness of the membrane cap 45 8 12 Testing the sensor electronics d...

Page 5: ...in this manual T042018 1 You must read and follow the sensor precautions before attempting to start up the analyzer to prevent damage to the sensor Please read this entire manual before unpacking sett...

Page 6: ...ature of the media due to an integrated temperature compensation pH has an effect on chlorine measurement accuracy since the sensor measures only the hypochlorous acid HOCl form of chlorine and not th...

Page 7: ...possible the sensors can be operated under a constant pressure of up to 1 bar relative pressure or 2 bar absolute pressure Fluctuations in pressure must be avoided The membrane cap must only be filled...

Page 8: ...pools NOTE Iron and manganese levels above USEPA Secondary Drinking Water Standards MCL will prevent the operation of the 3 electrode chlorine sensor The sensor is not suitable for detecting the absen...

Page 9: ...lorine including iso cyanuric acid chlorine independent of the iso cyanuric acid concentration Combined chlorine chloramine is not measured The sensor is not suitable for determining organic chlorinat...

Page 10: ...prevent the sensors from working correctly If the membrane comes into contact with tensides detergents surfactants from washing cleaning or disinfecting agents the hydrophobic properties of the membra...

Page 11: ...d nut 2 Terminal Cover 3 O ring 4 2 pin terminal for measuring cable connection 5 Electrode shaft with integrated electronics 6 Counter electrode stainless steel 7 O ring 8 Electrode finger reference...

Page 12: ...5 12 11 6 7 8 9 10 1 Cable gland nut 2 Terminal Cover 3 O ring 4 2 pin terminal for measuring cable connection 5 Electrode shaft with integrated electronics 6 O ring 7 Electrode finger reference elec...

Page 13: ...o point adjustment Not required pH value operating range 4 to 12pH pH dependency loss ofslope Linear decrease of approx 5 per eachupward pH unit starting from pH 7 Disruptive substances crosssensitivi...

Page 14: ...0 20 mg l 0 1 mg l for 50 100 200 mg l Response time t90 About 30 sec About 2 min Operating temperatures temperature compensation 5 to 45 C Zero point adjustment Not required pH value operating range...

Page 15: ...he flow cell part no 303500 is fastened to a wall or mounting panel with mounting bracket part no 303501 1 2 3 4 5 1 Sensor 2 Mounting bracket 303501 3 Connection G1 4 for hose 8mm x 6mm 4 Flow cell h...

Page 16: ...ing and installing the sensor 1 make certain the O rings and threads are clean and undamaged The transparent inspection glass 6 can be unscrewed from the fitting housing for cleaning 1 First push the...

Page 17: ...tainless steel membrane disc holder from the membrane cap this will cause the membrane to be misaligned and damage it The steps must be performed exactly as they are described below NOTE In order for...

Page 18: ...n installing the membrane cap to prevent electrolyte from leaking doing so will create excessive pressure that will destroy the membrane 6 Holding the shaft of the sensor vertically do not hold the me...

Page 19: ...through the membrane the membrane is faulty and you must use a new membrane cap 8 Take the stepped collar smaller diameter up and insert a screwdriver into the split to enlarge the opening 9 Push the...

Page 20: ...cell 13 Slide the cable gland nut and terminal cover over the sensor cable in the order shown The cable shrink wrap should protrude through the cable gland nut 14 There are two possible sensor cable...

Page 21: ...olution ground is necessary Sensor Conditioning Time The sensor will not provide a constant value until after a conditioning time of 2 hours has passed It cannot be calibrated until then The calibrati...

Page 22: ...with distilled water and dry 3 Using the special abrasive paper clean only the tip of the dried electrode finger To do this place the special abrasive paper on a paper towel with the matte side facin...

Page 23: ...ed with electrolyte onto the shaft of the sensor Storage transport You must store the sensor if it will not be used for monitoring for over 24 hours 1 Unscrew the membrane cap from the shaft of the se...

Page 24: ...or with SPST switch 2 meter detachable cord with IEC 60320 C13 NEMA 5 15P connectors Input Fuse 6 amp 125V fast acting 5 x 20 mm glass tube Designed to UL CSA 248 14 Electronics Microprocessor based T...

Page 25: ...Analyzer on Flow Panel 1 5 Amp fused power input module 2 24VDC power supply 88 264 Volts AC 50 60 Hz Input 1 Sample inlet 2 way ball valve 2 Flow Cell 3 Sensor 4 Flow meter with valve 12 GPH 5 Pressu...

Page 26: ...Wear an anti static wrist strap connected to an earth ground Component Identification screen display connection output is active Power supply under main circuit board not shown see previous page 1Amp...

Page 27: ...should be run with 3 8 diameter rigid PVC tubing to minimize lag time between the process and analyzer If a long sample line run is unavoidable increase the flow rate to the analyzer to about 10 ft s...

Page 28: ...ng feet included 2 3 Serial Number Software Version Number The FX CL F T chlorine analyzer is shipped with two serial number labels found on the bottom right side of the enclosure and on the inside of...

Page 29: ...the included 5 16 18 x 1 2 flat head screws 2 Anchor securely to the wall or panel The analyzer on mounting panel weighs 19 pounds Plumbing Connection 1 Connect the supplied 1 4 ID x 3 8 OD x 3 long...

Page 30: ...supply is required The analyzer is supplied with a 2 meter long power cord with IEC 60320 C13 and NEMA 5 15P connectors The power input module has an IEC male inlet to accept an IEC 60320 C13 connecto...

Page 31: ...ture input 13 Programming input connector factory use only 4 4 20mA outputs 1 active 14 Reset switch factory use only 5 Jumper Free Total chlorine sensor factory use only 15 Motor start stop 6 Chlorin...

Page 32: ...ensor Input 5 42 Display Wiring 5 43 Power Input 24V TB 2 Gnd TB 1 Flow Message Jumper Only Blue with white dots or Black TB9 24 White with blue dots or White TB9 1 Shield clear TB9 COM Blue TB2 4 Whi...

Page 33: ...UT TO AVOID GROUND LOOPS WHICH CAN DISRUPT THE MEASURING ELECTRODE DO NOT APPLY POWER TO THIS mA OUTPUT 5 45 RS485 Serial Output Wiring CAUTION CONNECT TO ISOLATED INPUT 5 46 Wiring Form C 1A 24VDC re...

Page 34: ...ature If the pH measurement option is included the sample pH will also be displayed The heartbeat indicator will flash in the lower left corner 4 Slowly open the sample inlet ball valve to deliver sam...

Page 35: ...are not possible until the sensor has become conditioned 9 Calibrate the chlorine sensor span standard using the concentration determined by a grab sample analysis using an approved method It is not n...

Page 36: ...y touching the screen Items within the black field are displays only of live online measurements and alarm conditions The large red values are the disinfectant concentration with unit of measurement b...

Page 37: ...Menu Structure MAIN RANGE Span Set ALARMS High Alarm Set High Alarm Delay Low Alarm Set Low Alarm Delay CALIBRATION Chlorine Zero Chlorine Standard SETTINGS Screen Settings Screen Brightness Volume pH...

Page 38: ...nalyzer Setting the range automatically scales the 4 20 mA output to the selected range and is not further scalable Setting the range also automatically sets the high and low alarm set points to 10 an...

Page 39: ...takes effect If needed reset the alarm levels from their default values 5 5 Configuring the mA Output CAUTION the current inputs of any external device connected to the analyzer current output must b...

Page 40: ...onfiguration screen The alarms will be set by default to 10 and 90 of range upon setting the operating range 2 To set the High residual alarm touch the Edit button On the High Alarm Set screen touch t...

Page 41: ...rometric titration due to various interferences and inherent limitations of the method The accuracy of the FX CL F T can never be greater than the instrument it is being calibrated against Accuracy is...

Page 42: ...he measuring cell when calibrating always take the sample from the flow cell drain or sample tap If the value for the nominal slope is under 30 the membrane cap and the electrolyte must be replaced an...

Page 43: ...rsion number and display version number These numbers are required for customer service and technical support 7 82 Screen Settings The screen brightness and audio volume can be adjusted to suit your n...

Page 44: ...n the membrane cap Replacemembrane cap Gas bubbles on the outside of the membrane Increase the flow briefly Check mountingand change ifnecessary No electrolyte in the Fill the membrane cap with electr...

Page 45: ...conditioned by the manufacturer Brown gray colors are normal 8 11Testing the leak tightness of the membrane cap 1 Carefully dry the outside of the membrane cap to be checked 2 Prepare the membrane cap...

Page 46: ...ectant 4 Move the sensor around in the beaker for approx 30 s without creating air bubbles 5 Leave the sensor in the beaker for 1 hour and wait for the settling time to elapse 6 Read the signal from t...

Page 47: ...ance schedule below is recommended Some applications may require more cell maintenance due to high levels of suspended solids high calcium levels or other characteristics of the sample liquid 9 1 Main...

Page 48: ...lyte for 2 electrode Free chlorine 100 ML 303300 Electrolyte for 3 electrode Free chlorine reduced ph dependence 100 ML 303301 Electrolyte for Total Chlorine 100 ML 303320 Sensor Free chlorine 2 elect...

Page 49: ...placement must be freight prepaid and include the following information 1 A clearly written description of the malfunction 2 Name of person to contact and the phone number where they can be reached 3...

Page 50: ...y unauthorized repair Buyer agrees to hold Foxcroft Equipment Service harmless from all claims for damages arising out of injury or death to any person or damage to any facility or any other property...

Page 51: ...51...

Page 52: ...ervice Co Inc 2101 Creek Road P O Box 39 Glenmoore PA 19343 Tel 800 874 0590 610 942 2888 Fax 610 942 2769 Email service foxcroft com www foxcroft com Serial Number Label for FX CL Amperometric Chlori...

Reviews: