29
Section 4.5 Wall Mounting, Plumbing
Wall mounting the analyzer
Locate at approximate shoulder height for convenience. The FX-CL-F/T should be level from side to
side and front to back.
1.
Attach the four mounting feet to the rear of the electronics enclosure using the included 5/16"-
18 x 1/2" flat head screws.
2.
Anchor securely to the wall or panel. The analyzer on mounting panel weighs 19 pounds
Plumbing Connection
1. Connect the supplied 1/4” ID x 3/8” OD x 3’ long, clear flexible PVC sample hose to the
existing plumbing. Cut if necessary to keep the sample line as short as possible.
2. Route the supplied 1/4” ID x 3/8” OD x 3’ long, clear flexible drain tubing as needed.
3. If needed cut the grab sample tap tubing to desired length.