Section 1.13 Sensor & Flow Cell Assembly
Overview
(1)
(2)
(8)
(9)
(10)
(3)
(4)
(7)
(5)
(6)
(1)
Sensor
(6)
Inspection glass
(2) Union nut
(7)
Inlet G1/4A or DN10
(3) Flow Cell
(8)
Stepped collar 1"
1
(4) Outlet G1/4A or DN10
(9)
Compression ring
1
(5) O-ring, inspection glass
(10) O-ring, stepped
collar
1
1
Part of the flow cell.
Caution:
When assembling and installing the sensor (1) make certain the O- rings and
threads are clean and undamaged
The transparent inspection glass (6) can be unscrewed from the fitting housing for cleaning.
1.
First push the O-ring (10) onto the sensor (1), then the compression ring (9)
and over it the 1" stepped collar (8) (starting from the Pg screw connection).
The stepped collar (8) must engage in the groove.
2.
After the sensor has been prepared in this manner, insert it into the flow cell
(3) and fasten it in place with the union nut (2).