F1 F2 F3 F4
Press
YES (F1)
key to verify the Zero CAL-CHECK™.
F1 F2 F3 F4
WARNING: You must verify that there is a no flow condition before
proceeding. Be sure to isolate the sensor completely - any air movement
(even from a fan) can result in a false "Fail" result.
Once process is stable, press
YES (F4)
key to begin the Zero CAL-CHECK™.
F1 F2 F3 F4
This test will take less than 5 minutes. The T=xx is a count down timer
indicating how much time is left to finish the test.
F1 F2 F3 F4
Upon test completion, the final value will be displayed along with a Pass/Fail
message.
Diff=x.xxx %
Passed OK
Process Zero and
Stable? YES
Verifying ZRO
Diff = xx.xx T=xx
VERIFY ZRO?
YES NO
22
Operation: Zero CAL-CHECK
™
Zero CAL-CHECK
™
-
Out of Pipe
Calibration Validation