i
Performing the CAL-V
™
Calibration Validation Test
This menu allows the user to confirm the calibration of the FT3 by verifying
the functionality of the sensor and sensor signal processing circuitry. During
the CAL-V™ calibration validation test, the microprocessor adjusts current to
the sensor elements and determines the resulting electrical characteristics.
These site characteristics are compared with the data that was collected
at the factory during the original meter calibration. Matching data within
established tolerances confirms the meter is accurate. This test can be
performed under no flow or normal flow conditions. The test takes up to four
minutes to complete. At the conclusion of the test, a Pass or Fail message
will be displayed. Press
F4
at the conclusion of the test to return to normal
measuring mode or to terminate the test.
Press
FLO (F2)
from the main menu. The display will show:
F1 F2 F3 F4
Press
DGN (F1).
The display will show:
F1 F2 F3 F4
Press
TST (F2).
The display will show:
F1 F2 F3 F4
Press
CALV (F1)
to perform the CAL-V™ verification test.
Note: The FT3 will stop measuring flow when performing this test.
Press
EXIT (F4)
to exit if you do not wish to continue.
FLOW PARAMETER 1
DGN UNT PRM EXIT
DIAGNOSTIC
SIM TST EXIT
16
Operation: CAL-V
™
CAL-V
™
Calibration Validation
OPERA
TION
DIAGNOSTIC
CALV ZRO EXIT