Large Nut
Small Nut
End Fitting
Adjacent Hex
Fittings
7/8
” w
ren
ch
13/
16
” w
ren
ch
Small Nut
Adjacent Hex
Fig. 2.3: Close-Up: Fittings of Fox Packing Gland Assembly
In order to isolate the sensor, it must be retracted from the pipe. Loosen the
small nut of the compression fitting using 7/8" and 13/16" wrenches. Once
loosened, the probe will be able to slide upwards (see Figure 2.4 below).
Fig. 2.4: Loosening the Compression Fitting
11
Start Up
Calibration Validation
ST
ART UP