Figure 1.3.3. The cassette with groove plate
CHAPTER 2
2.1 Installation place and environmental conditions
It is strongly recommended to pay attention to where the heat oven will be
installed. The installation place should satisfy the following conditions.
Temperature:
10
to 40°C
Relative humidity:
15 to 85% RH (non condensing)
Storage temperature:
0 to 50°C
Furthermore, the unit should not be in any of the following kind of places:
-
places exposed to direct sunlight
-
places with violent temperature swings
-
places frequently exposed to mechanical vibrations
-
places that are very dusty
-
places that are poorly ventilated
Do not install the oven in places that are near an air conditioning vent.
2.2 Power supply
The power supply conditions for the OMC-W36 heat oven are as follows:
Model OMC-W36-1:
•
Power supply voltage .................. single phase AC 110V ± 10%
•
Power supply frequency .............. 50/60 Hz ± 10%
•
Power consumption rating ........... 250W
Summary of Contents for OMC-W36