39
The meter should be replaced after the validated number of
cleaning and disinfection cycles or the warranty period, which
ever comes first.
Stop using the meter if you see any signs of deterioration, such as
Meter can not be turned on
LCD display cracks or becomes cloudy
buttons no longer function
outer casing cracks
Data can not be transmitted to PC
Inaccurate results are obtained
Please contact the local customer service for a replacement
meter if any of the signs of deterioration are noticed.
Note
Do NOT clean and disinfect the meter while performing tests.
Do NOT allow cleaning and disinfecting solution to get in the
test slot, battery compartment, or strip-ejection button.
If you do get moisture in the test strip slot, wipe it away with a
corner of tissue.
Always dry the meter thoroughly before using it.
Do not spray the meter directly with cleaning solutions espe-
cially those containing water (i.e. soapy water), as this could
cause the solution to enter the case inside and damage the
electronic components or circuitry.