38
Disinfecting Procedures:
1.) Put on non-sterile gloves.
2.) Take out one disinfecting wipe from the package and squeeze
out any excess liquid in order to prevent damage to the meter.
3.) Wipe all meter’s exterior surface display and buttons. Hold the
meter with the test strip slot pointing down and wipe the area
around the test slot but be careful not to allow excess liquid to
get inside. Keep meter wet with disinfection solution contained
in the wipe for a minimum of 2 minutes for Micro-Kill+™ wipes.
Follow the instructions on the package label of disinfecting
wipe. Use two or more wipes if necessary.
4.) Remove the wipe. Allow the meter surface to dry completely.
5.) Discard the used wipes and never reuse them.
6.) Remove and discard gloves in appropriate receptacles and
wash hands.
Improper system cleaning and disinfection may result in meter
malfunction. If you have a question, please contact the customer
service for assistance.
This device has been validated to withstand up to 10,000 clean-
ing and disinfection cycles using the recommended disinfecting
wipes/towelettes. The tested number of cycles is estimated by
approximately 10 cleaning and disinfection cycles per meter per
day for 3 years, the expected life of the device.