12
1. Insert the test strip to turn on the
meter
Insert a test strip into the meter. Wait
for the meter to display “
” and “ ”.
2. Press M button to mark this test as
a control solution test
WARNING:
When doing the control solution test, you have to mark it so that the test
result will
NOT
mix with the blood glucose
TEST RESULTS
stored in the
memory. Failure to do so will mix up the blood glucose test results with the
control solution test results in memory.
3. Apply control solution
Shake the control solution vial thoroughly before use. Squeeze out a
drop and wipe it off, and then squeeze out another drop and place it on
the tip of the vial cap. Hold the meter in an angled position so that the
absorbent hole of the test strip touches the drop. Once the confirmation
window fills completely, the meter will begin counting down. To avoid
contaminating the control solution, do not directly apply control solution
onto the strip.
Performing a Control Solution Test