background image

22

23

For setting the orientation of this angle the complete rotation head upper and lower half 

should be turned together to the desired orientation. If the setting is done fixate the rotation 

head by tighten the bolt with 6 Nm.

Installing the chair / stand mount

When a Dowing

2

 needs to be mounted on a working chair or stand the Dowing

2

 should be 

mounted on the chair/stand mount. 

Setscrews to fixate the Chair/ Stand mount

The chair/stand mount is designed to fit on a square tube 20 x 20 x 1.5 mm.

To install the chair/stand mount slide it over the tube and tighten the screws. For a proper 

installation, the tube should slide at least 25 mm into the bracket. Tighten both screws with 

6 Nm.

Adjusting the angle of the rotation head

Dowing

2

 will perform the best when the rotation head is leveled. By default the rotation head 

is parallel with the table- and chair/ stand mount (i.e. the stripes on the two rings are in line). 

In some cases it is desired to adjust the angle of the rotation head. For example when a table is 

tilted, or according to user needs.

Bolt to fixate 

the roation head

Lower part

Upper part

To change the angle of the rotation head, first the bolt which fixates it has to be loosened. 

Second, only the upper half of the rotation head should be turned to the desired angle. 

Summary of Contents for Dowing2

Page 1: ...ID manual 606281 Revision 1 0 0 Date 1 May 2020 User manual Dowing2...

Page 2: ...trademark of Focal Meditech BV Dowing2 is a registered trademark of Focal Meditech BV Dowing2 Model Design is registered by Focal Meditech BV Content Disclaimer 2 Copyright Notice 2 Trademarks 2 Cont...

Page 3: ...essential information for using Dowing2 information about safety issues and contact information Please read this information carefully increase of knowledge of the arm support will result in an increa...

Page 4: ...aution This symbol in combination with the word Caution is used to warn about possible unsafe practices Extra attention is required Disposal This symbol indicates that this product is not to be dispos...

Page 5: ...ical energy is stored for balancing the arm During non intended removal of the arm from the arm fitting of Dowing2 this mechanical energy will be released resulting in a fast moving human interface wh...

Page 6: ...aluation attention is required to determine if the user is able to sit in a stable position and if one can stand up without using a support Intended use and operation of the device Intended use Dowing...

Page 7: ...all intended to be used by persons challenged by limited muscle force in their arms and shoulder girdle Due to diminished use of their musculoskeletal functions prior to the supply of Dowing2 and als...

Page 8: ...ontaining the loadarm arm fitting elbow fitting and wrist support Body of Dowing2 The body of Dowing2 contains the balancing system including the possibility to adjust the balancing force Balancing fo...

Page 9: ...itting STEP 2 Apply a small force downwards and slide the locking ring forwards over the hook Disconnecting the human interface Danger Please carefully disconnect remove the different parts and take n...

Page 10: ...g2 STEP 2 Remove the user s arm STEP 3 Remove Dowing2 from the parking position Warning the arm fitting will move upwards due to the energy stored in the spring STEP 4 Remove the human interface See d...

Page 11: ...r the user to make a straight line To change the angle loosen the bolt first before adjust the locking hook If the locking hook is set to the desired angle tighten the bolt with 6 Nm Edge depth Edge d...

Page 12: ...over the tube and tighten the screws For a proper installation the tube should slide at least 25 mm into the bracket Tighten both screws with 6 Nm Adjusting the angle of the rotation head Dowing2 will...

Page 13: ...lifetime of system parts and interfere with operation All housings must be regularly inspected If any housing is visibly damaged do not use the device It is prohibited to physical modify Dowing2 Ther...

Page 14: ...0 45 45 Dowing2 dimensions human interface excluded Dimensions of the Dowing2 are variable This is because it is a device with moving parts and different configurations All dimensions are given in mm...

Page 15: ...03145 Elbow fitting right X Large 603140 Elbow fitting left XX Large 603146 Elbow fitting right XX Large 603109 Arm fitting size 1 left 603111 Arm fitting size 1 right 603113 Arm fitting size 2 left 6...

Page 16: ...ght Small 603135 Elbow fitting left Large 603141 Elbow fitting right Large 603139 Elbow fitting left X Large 603145 Elbow fitting right X Large 603140 Elbow fitting left XX Large 603146 Elbow fitting...

Page 17: ...ize 5 right 605658 Inlay arm fitting size 1 left Neoprene 605638 Inlay arm fitting size 1 right 605662 Inlay arm fitting size 2 left 605642 Inlay arm fitting size 2 right 605666 Inlay arm fitting size...

Page 18: ...ick start guide contains the basic information regarding Dowing2 The aim of this information is to ensure successful safe and effective use of the device This quick start guide will give you a quick l...

Page 19: ...leiden tot onjuist functioneren van het apparaat of tot lichamelijke schade Danger Prevent extreme temperature see environment conditions in user manual Failure of this can lead to malfunctioning of...

Page 20: ...t using Dowing2 Parkeer of start gebruik Dowing2 If Dowing2 is not used for a short time it is recommended to place the Dowing2 in its intended parking bracket This will provide a safe situation for b...

Page 21: ...tkoppelen van de Human Interface Figure 4 Attach detach human interface Figuur 4 Koppelen ontkoppelen human interface The human interface including the loadarm arm fitting elbow fitting and wrist supp...

Page 22: ...one hand the body of Dowing2 with the other hand lift up in a straight line the lower swivel arm from the clamp STEP 6 Store Dowing2 in its case Dowing2 loskoppelen STAP 1 Vergrendel Dowing2 in de ve...

Page 23: ...a 31 0 13 533 31 03 en we zenden u een geprint exemplaar toe 12 Adjust the balancing force Aanpassen balanceerkracht Figure 7 Adjust the balancing force Figuur 7 Aanpassen balanceerkracht The upwards...

Page 24: ...46 47 Appendix 5 Declaration of conformity 14...

Page 25: ...Focal Meditech are subject to conditions of the national or local representative and in accordance with national law Conditions and Warranty direct supply by Focal Meditech BV to consumers In the case...

Page 26: ...50...

Reviews: