Adjusting the Thermal Image
19
Fast Auto/Manual Range Toggle
When NOT in a menu mode, press
c
for ½ second and release to toggle
between Auto Range and Manual Range.
Fast Auto Rescale
When in Manual Range and NOT in a menu mode, press
c
for ½ second and
release to automatically rescale the level and span range for objects within the
Imager’s thermal field of view.
Note
The Imager always powers up in the same Range mode, Auto or
Manual, as when it powers down.
Setting the Level (Manual)
When in manual ranging, the level setting adjusts the mid-point thermal span
within the total temperature range of the Imager. To set the level:
1.
After entering the manual range mode (see “Setting the Range”), press the
softkey labeled
Go to Level
. This will place the Imager into the
Adjusting Level
mode.
2.
Press the softkey labeled
Up
to move the temperature span to higher
temperatures or
Down
to move the span to lower temperatures.
3.
To adjust the Span, press the softkey labeled
Go to Span
(see “Setting
the Temperature Span”).
4.
To capture an image, squeeze and release the trigger once. See Figure 2.