background image

T+/T+PRO English Instruction Sheet 

Page  

 

Optional Accessories 

Accessories are available for the T+/T+ PRO Electrical Tester. 
See Table 2. 

Table 2. Accessories 

Item Item 

Number 

Belt Holster  

H3 

Replaceable Test Probes (1 red and 1 black) 

TP2 

Replaceable Test Lead assembly without probes 

2792344  

Pushbuttons 

Use the pushbuttons to toggle switchable functions of the Tester 
on or off. The pushbuttons are shown and described in Table 3. 

Table 3. Pushbuttons 

Pushbutton Description 

C

 

Press to turn the flashlight on, release to turn it off.  
The flashlight will turn off 5 seconds after release of the 
pushbutton. Batteries are required for this feature. 

O

 

Toggles the ohms function on or off. T+PRO only. See 
“Measuring Resistance (T+ PRO Only)”. 
Batteries are required for this feature. 

H

 

Toggles the HOLD function on or off. T+PRO only.  See 
“HOLD”. Batteries are required for this feature. 

Z

 

Turns beeper on or off. When the beeper is turned off, it 
will not sound until it is reactivated. Batteries are 
required for this feature. 

G

 

Performs Ground Fault and trips the GFCI. 

1.800.868.7495

[email protected]

Fluke-Direct

.ca

Summary of Contents for T+

Page 1: ...O only Flashlight GFCI trip Additional languages of this instruction sheet Brazilian Portuguese Simplified Chinese Japanese Korean and Thai are available for download at www fluke com Contacting Fluke...

Page 2: ...t operate the Tester around explosive gas vapor or dust Do not apply more than the rated voltage as marked on the Tester between terminals or between any terminal and earth ground When measuring hazar...

Page 3: ...ge source for a MAXIMUM of 30s and then disconnect for a MINIMUM of 300s Perform a self test before any measurements are taken for voltage continuity resistance or rotary field See Self Test Exercise...

Page 4: ...ose of this product as unsorted municipal waste Contact Fluke or a qualified recycler for disposal N10140 Conforms to relevant Australian standards CAT III CAT III equipment is designed to protect aga...

Page 5: ...n C Press to turn the flashlight on release to turn it off The flashlight will turn off 5 seconds after release of the pushbutton Batteries are required for this feature O Toggles the ohms function on...

Page 6: ...urements cannot be made and the LCD will not switch on WCaution Replace depleted batteries immediately to avoid damage to the Tester from battery acid leakage B All audible signals are disabled C HOLD...

Page 7: ...ed for this feature Vibration Voltage Indication The Tester vibrates when voltages 40 V and higher are measured Batteries are required for vibration to function Automatic Operation The Tester automati...

Page 8: ...y Perform a self test before any measurements are taken for voltage continuity resistance or rotary field When the batteries are depleted self test will not function Replace the batteries WCaution Alt...

Page 9: ...T PRO will show the measurement on the LCD for voltages 10 V 10 V the LCD remains blank The AC DC and DC LEDs will only operate for voltages greater than 24 V Refer to Figure 2 DC Voltage 600 V dc max...

Page 10: ...rcuit power and connect the probes to the circuit under test Continuity a resistance less than 20 k is indicated by the beeper and a lighted P If the beeper is disabled it will not sound during this t...

Page 11: ...are discharged To measure resistance turn off the circuit power press O and place the test probes across the resistance under test For resistances less than 9 99 k the T PRO display shows the resistan...

Page 12: ...ontinue to indicate the presence of voltage When testing and verifying GFCI circuits or components do the following Consult the GFCI manufacturer s installation instructions to determine that the GFCI...

Page 13: ...erating properly Rotary Field Direction T PRO Only This test shows the direction of a rotary field in a three phase system when measured phase to phase The LCD indicates phase rotation for a right or...

Page 14: ...g for an indication of a change in the field direction CONTIN UITY 120 208 48 24 12 240 277 600 480 CONTIN UITY 120 208 48 24 12 240 277 600 480 600 CONTINUITY 120 208 48 24 12 240 277 480 eqc05f emf...

Page 15: ...the battery compartment door and reinstall the screw Specifications Temperature Operating 10 C to 55 C 14 F to 131 F Storage 30 C to 60 C 22 F to 140 F Relative Humidity 0 C to 30 C 32 F to 86 F 90 30...

Page 16: ...tteries or damage from accident neglect misuse alteration contamination or abnormal conditions of operation or handling Resellers are not authorized to extend any other warranty on Fluke s behalf To o...

Reviews: