SONOCHEK
Technical data
108
SONOCHEK 09.2018
Saving and
recording
Audio data: Format: *.wav
Storage location: Internal measurement memory
Memory requirement
Per second of measurement time
Level
1 kB
Spectra
31 kB
Audio
16 kB
Displays
-
Level wheel and level bar (only current level)
L
– current level table
(in dB, reference p
0
= 20 µPa, temporal resolution: 16 ms per level value)
-
Spectrogram (temporal resolution: 16 ms per frequency spectrum)
Operating modes
Live
–
display of the current readings,
Recording
–
display and storage of the current readings
Languages
German, English, Spanish, French, Portuguese, Turkish, Italian, Chinese,
Dutch, Russian, Polish, Japanese
AMBIENT CONDITIONS
Operating temperature
-10 °C
…
+40 °C
Temperature when
charging the battery
0 °C ... +20 °C
Storage temperature
-20 °C
…
+60 °C
Degree of protection
IP40
Standards and
guidelines
EMC directive 2014/30/EU; WEEE [Waste Electrical and Electronic
Equipment]directive 2012/19/EU; RoHS [Restriction of Hazardous
Substances Directive] directive 2011/65/EU; ASTM E1002-2005
Accessories (included
in the scope of delivery)
-
Sensor system: Airborne sound sensor DBS10 with different attachments,
structure-borne sound sensor DBS20 with integrated temperature sensor
and different probe tips
-
Software: Measurement software "SONOLEVEL app" for basic
maintenance activities, application software "SONOLEAK app" for the leak
detection and filtering
-
Hearing protection headphones, device case, charger, protective sleeve,
protective film, carry strap