RUSKA 7350
Users Manual
A-2
The key to eliminating an error is knowing its source and type along with its polarity and
magnitude. Generally, the source is simple to detect and is represented by the specific
parameter. The type is usually a function of the instrument’s design and manufacturing
process. Within a given instrument, an error can be either random or systematic, as well
as random or systematic within the instrument’s population. The user is free to consult
Fluke for recommended methods of minimizing error source contributions.
In summation, total error can and should be managed by the control of the three general
error sources: Input Specifications, which includes the user’s chosen calibration standards;
General Specifications, which includes the user’s chosen processes; and Performance
Specifications, which includes the user’s chosen applications for the instrumentation. The
parameters and value limits listed in the following specifications indicate the product
line’s general acceptance limits and are not a report of any unit’s specific error
contribution. Any parameter exceeding the specified limits should be considered in need
of maintenance.
Specifications
Pressure Ranges
0-6000, 0-10,000 psia
Accuracy
0.01% FS
Resolution
0.001% FS or better
Control Stability
Typically 0.0075% FS
Control Response
20 – 90 seconds controlling into a 15 cu. in. volume
depending on full scale range and increment step size
Control Range
300 – 3000, 600 – 6000, 1000 – 10,000 psia
Control Volume
3 in
3
(50cc) to 35 in
3
(570cc) all ranges
Overshoot
0.01% of Full Scale
Overpressure Protection
Relief valve set at 110% FS
Connections
Supply Pressure port: 1/8
NPT
Test port: 1/8 NPT
Drive Air Supply: 1/4 NPT
Display
Color active matrix TFT display
Pressure Medium
(test and control gas)
Clean, dry, non-corrosive gas with a dew point of -40 °C
and a particulate size
<
100 microns.
Gas Consumption
Zero at commanded pressure with a leak tight system
Pressure Supply
110 to 120 psi gas pressure for controller circuit. Test gas
supply to be 110% higher than full scale pressure (may
require an external booster system).
Recommended
Recalibration Interval
1 year
Communications
Standard: RS-232 and IEEE-488
Dimensions
19" W x 24" D x 17" H
(49 x 61 x 43 cm)
Weight 120
lbs
Summary of Contents for RUSKA 7350
Page 8: ...RUSKA 7350 Users Manual vi ...
Page 10: ...RUSKA 7350 Users Manual viii ...
Page 16: ...RUSKA 7350 Users Manual 1 6 ...
Page 22: ...RUSKA 7350 Users Manual 2 6 ...
Page 52: ...RUSKA 7350 Users Manual 4 28 ...
Page 62: ...RUSKA 7350 Users Manual 5 10 ...
Page 72: ...7350 Users Manual 7 4 ...
Page 76: ...RUSKA 7350 Users Manual A 4 ...