RUSKA 7350
Users Manual
B-2
Value
Description and Corrective Action
-286
Program Runtime Error. An error occurred while running the program. Usually the
setpoint is out of range.
-313
Calibration Data Lost. The calibration data has been lost and the unit must be
recalibrated.
-315
Configuration Data Lost. The configuration data has been lost. Check all parameters
to be sure that they are correct.
-330
Self Test failed. See display for more information.
-350
Queue Overflow. The error queue was full and messages were lost.
-400
Query Error. A read request was received when there was nothing to read.
500
Controller Malfunction. Internal control failure.
501
High Limit Exceeded. The pressure was greater than the high limit.
502
Low Limit Exceeded. The pressure was less than the low limit.
503
Slew Limit Exceeded. The pressure changed faster than the slew limit allowed.
521
Pressure Overrange. The pressure reading is outside the range of the RUSKA 7350.
538
Auto vent. The Vent Limit was exceeded and the instrument entered Vent Mode.
543
Supply Pressure. The instrument is attempting to increase pressure and the
pressure is not responding. Check main pressure supply, shop air supply, and for
leaks.
600
Factory Data Lost. Internal factory constants have been lost. Contact Fluke for more
information.
602
Model Changed. Firmware has been downloaded that is not correct. Download the
correct firmware.
706
Pump position failure. The position sensor is not detecting movement in the pump
after the pump has been commanded to move.
707
Range Switch. The current range can only be changed when the pressure reading is
below 100 psi (700 kPa).
Summary of Contents for RUSKA 7350
Page 8: ...RUSKA 7350 Users Manual vi ...
Page 10: ...RUSKA 7350 Users Manual viii ...
Page 16: ...RUSKA 7350 Users Manual 1 6 ...
Page 22: ...RUSKA 7350 Users Manual 2 6 ...
Page 52: ...RUSKA 7350 Users Manual 4 28 ...
Page 62: ...RUSKA 7350 Users Manual 5 10 ...
Page 72: ...7350 Users Manual 7 4 ...
Page 76: ...RUSKA 7350 Users Manual A 4 ...