Power Analyzer
Specifications
83
Accuracy
Accuracy
(% r % range)
Range
DC
AC
10 Hz to 1 kHz
10 kHz
Shunt
10 A
0.1 + 0.2
0.1 + 0.1
5 + 1
1 A
0.1 + 0.5
0.1 + 0.2
5 + 1
0.1 A
0.1+2
0.1 + 1
5 + 1
BNC
10 V
0.1 + 0.1
0.1 + 0.1
5 + 1
1 V
0.1 + 0.2
0.1 + 0.1
5 + 1
0.1 V
0.1 + 1
0.1 + 0.5
5 + 1
Motor (NORMA 6003+
、
NORMA 6004+)
Voltage Range ................................... ±10 V dc, 10 % overload
Voltage Channels ............................ 2
Input Impedance ................................
1.1 MΩ (Typical)
Temperature Coefficient .................. 0.05 x (Spec)/K
Accuracy @ DC ............................... 0.1 % r 0.1 % range
Pulse Channels ................................. 3
Pulse logic high threshold ................ 2 V (Typical)
Pulse logic low threshold ................. 0.8 V (Typical)
Maximum pulse frequency ............... 100 kHz
Frequency Domain Measurement
Frequency Accuracy .......................... 0.05 % range + 0.05 % reading
(3 ranges: 1 Hz to 10 Hz, 10 Hz to 400 Hz, 400 Hz to 100 kHz)
Harmonics ......................................... 100 order, up to 6 kHz
Summary of Contents for NORMA 6003
Page 6: ......
Page 98: ...NORMA 6003 NORMA 6003 NORMA 6004 NORMA 6004 User Manual 92 ...