NORMA 6003/NORMA 6003+/
NORMA 6004/NORMA 6004+
User Manual
54
Item
Description
Move
C
Move the cursor
Each time you push
C
, the currently selected cursor will be cycled through.
The selected cursor is indicated in yellow and is displayed below the “
Move
”
label.
After selecting a cursor, use the arrow (
M N O P
) keys to move the cursor to
the corresponding position on the waveform.
Cursor
D
Turn the cursor on/off
Push
D
to turn the cursor function on or off.
Use the Cursor
The following content uses the double vertical lines as an example to introduce the use of the cursor,
which is similar to the use of the cursor of a single vertical lines and double horizontal lines.
1. From the main screen of Scope select
D
(
Cursor
).
2. Push
D
(
Cursor
) until “ON” is shows on the Softkey label.
3. Push
B
(
Type
) until
shows on the Softkey label.
4. Push
A
(
Select
) to select the signal to be measured.
The difference between the crossing points of the two cursors and the signal under measurement is
shown in the upper left corner of the screen, such as “204.7V”; the time difference (426ms) between
the two cursors and the frequency (2.347 Hz) calculated are displayed in the upper part of the screen.
5. Push
C
(
Move
) until
shows on the Softkey label, indicating that the left cursor is selected
and a diamond mark is displayed on the left cursor.
6. Use the left and right arrow (
O P
) keys to move the left cursor to the appropriate position.
7. Push
C
(
Move
) until
shows on the Softkey label, indicating that the right cursor is
selected and a diamond mark is displayed on the right cursor.
8. Use the left and right arrow (
O P
) keys to move the right cursor to the appropriate position.
9. Read the difference, time difference and frequency value at the top of the screen.
Summary of Contents for NORMA 6003
Page 6: ......
Page 98: ...NORMA 6003 NORMA 6003 NORMA 6004 NORMA 6004 User Manual 92 ...