1523, 1524 Reference Thermometer
Manual Calibration Process
86
39. Turn off T1 by sending the following command.
&$/'(95$1*(6(/1250$/
'LVFRQQHFWUHVLVWRU
40. Connect a 100 Ohm resistor to the calibration cable connected to the UUT on
Channel T2.
41. Query the reading every 2 seconds for 40 readings using the following
command:
0($6"
5HFRUGWKHGDWD
42. Turn off Channel T2 by sending the serial command:
&$/'(95$1*6(/1250$/
&RQQHFWWKHFDOLEUDWLRQFDEOHWR&KDQQHO7
7.9.1.3.3
Medium Ohms Range
43. Set the UUT in the MED_OHMS cal range by sending the following commands
through the serial port:
&$/'(95$1*6(/0('B2+06
44. Connect a 200 ohm resistor to the calibration cable connected to the UUT.
45. Query the reading every 2 seconds for 40 readings using the following
command:
0($6"
5HFRUGWKHGDWD
46. Connect a 400 Ohm resistor to the calibration cable connected to the UUT.
47. Query the reading every 2 seconds for 40 readings using the following
command:
0($6"
5HFRUGWKHGDWD
48. Connect a 10K Ohm resistor to the calibration cable connected to the UUT.
49. Query the reading every 2 seconds for 40 readings using the following
command:
0($6"
5HFRUGWKHGDWD
50. Connect a 40K Ohm resistor to the calibration cable connected to the UUT.
51. Query the reading every 2 seconds for 40 readings using the following
command:
0($6"
5HFRUGWKHGDWD
Note:
Steps 52 through 56 apply to the 1524 only. Proceed to step 57 if using a 1523.
Summary of Contents for Hart Scientific 1523
Page 14: ......
Page 17: ......
Page 19: ...11 Quick Start Setup Figure 6 1523 Menu Figure 7 1523 Menu cont ...
Page 22: ...1523 1524 Reference Thermometer Setup 14 Figure 9 1524 Menu Figure 10 1524 Menu cont ...
Page 23: ...15 Quick Start Setup Figure 11 1524 Menu cont ...
Page 24: ...1523 1524 Reference Thermometer Setup 16 Figure 12 1524 Menu cont ...
Page 35: ...27 General Operation INFO CON Connector Figure 16 Probe wiring diagrams ...
Page 36: ......
Page 50: ......
Page 102: ......
Page 104: ......
Page 106: ......
Page 110: ......