1523, 1524 Reference Thermometer
Specifications
18
Table 6
General Specifications
Operating
Temperature
†
–10 °C to 60 °C
0 °C to 60 °C with ac adapter
Storage Temperature
–20 °C to 70 °C
Operating altitude
10,000 meters above mean sea level
(2,000 meters with ac Adapter)
Relative Humidity (%
RH operating without
condensation)
0 % to 90 % (non condensing)
Vibration
Random, 2g, 5–500 Hz
Power requirements
3 AA alkaline batteries 12 V dc universal
power supply
Size
96 x 200 x 47 mm (3.75 x 7.9 x 1.86
inches)
Weight
0.65 kg (1.4 lb)
Safety
EN 61010-1:2001, CAN/CSA C22.2 No.
61010.1-04
†
Environmental conditions for all specifications: 13 °C to 33 °C
Table 7
Millivolt Measurement
Range Resolution
Accuracy
–10 mV to 75 mV
0.001 mV
± (0.005 % + 5 μV)
Temperature Coefficient ( –10 °C to 13 °C, +33 °C to 60 °C):
± (0.001 %/°C + 1 μV/°C)
Table 8
Reference Junction Compensation
Thermocouple Reference Junction Compensation
Accuracy
± 0.20 C
Table 9
Ohms Measurement, RTDs
Ohms Range
Accuracy ± Ω 4 Wire
0 Ω to 400 Ω
± (0.004 % + 0.002 Ω)
Temperature Coefficient ( –10 °C to 13 °C, +33 °C to 60 °C):
0.0008 %/°C + 0.0004 Ω
Excitation Current: 1 mA
Summary of Contents for Hart Scientific 1523
Page 14: ......
Page 17: ......
Page 19: ...11 Quick Start Setup Figure 6 1523 Menu Figure 7 1523 Menu cont ...
Page 22: ...1523 1524 Reference Thermometer Setup 14 Figure 9 1524 Menu Figure 10 1524 Menu cont ...
Page 23: ...15 Quick Start Setup Figure 11 1524 Menu cont ...
Page 24: ...1523 1524 Reference Thermometer Setup 16 Figure 12 1524 Menu cont ...
Page 35: ...27 General Operation INFO CON Connector Figure 16 Probe wiring diagrams ...
Page 36: ......
Page 50: ......
Page 102: ......
Page 104: ......
Page 106: ......
Page 110: ......