1523, 1524 Reference Thermometer
Specifications
20
Table 12
Temperature, Thermocouples External Reference Junction
Type
Range
Measure Accuracies
(ITS-90)
B
600 to 800 °C
0.85 °C
800 to 1000 °C
0.68 °C
1000 to 1800 °C
0.57 °C
C
100 to 550 °C
0.32 °C
550 to 2300 °C
0.71 °C
E
-200 to 0 °C
0.18 °C
0 to 950 °C
0.09 °C
J
-200 to 0 °C
0.21 °C
0 to 1200 °C
0.12 °C
K
-200 to 0 °C
0.31 °C
0 to 1370 °C
0.15 °C
L
-200 to 0 °C
0.15 °C
0 to 900 °C
0.11 °C
M
-20 to 0 °C
0.14 °C
0 to 400 °C
0.14 °C
400 to 1400 °C
0.14 °C
N
-200 to 0 °C
0.48 °C
0 to 1300 °C
0.20 °C
R
-20 to 0 °C
1.06 °C
0 to 500 °C
0.95 °C
500 to 1750 °C
0.48 °C
S
-20 to 0 °C
1.03 °C
0 to 500 °C
0.93 °C
500 to 1750 °C
0.55 °C
T
-200 to 0 °C
0.30 °C
0 to 400 °C
0.13 °C
U
-200 to 0 °C
0.27 °C
0 to 400 °C
0.13 °C
Resolution:
0.01 °C
Note 1:
Accuracies are based on external reference junction
compensation.
Summary of Contents for Hart Scientific 1523
Page 14: ......
Page 17: ......
Page 19: ...11 Quick Start Setup Figure 6 1523 Menu Figure 7 1523 Menu cont ...
Page 22: ...1523 1524 Reference Thermometer Setup 14 Figure 9 1524 Menu Figure 10 1524 Menu cont ...
Page 23: ...15 Quick Start Setup Figure 11 1524 Menu cont ...
Page 24: ...1523 1524 Reference Thermometer Setup 16 Figure 12 1524 Menu cont ...
Page 35: ...27 General Operation INFO CON Connector Figure 16 Probe wiring diagrams ...
Page 36: ......
Page 50: ......
Page 102: ......
Page 104: ......
Page 106: ......
Page 110: ......